PERFORM Operating Document 

 

Use and Cleaning Procedures for Bodystat

 

QuadScan 4000 

 

PC-POD-CP-012-v01 

 

 

Revision History 

Version 

Reason for Revision 

Date 

01 

New POD 

June 18, 2020 

 

 

1.

 

Introduction  

 

The content of this PERFORM Operating Document (POD) provides guidelines for use 
and the cleaning procedure for the Bodystat

®

 QuadScan 4000 multi-frequency bio-

impedance analysis unit. This unit assesses body composition measuring at 5, 50, 100, 
and 200 kHz. This POD applies to all users of the unit. 

Содержание QuadScan 4000

Страница 1: ...Revision History Version Reason for Revision Date 01 New POD June 18 2020 1 Introduction The content of this PERFORM Operating Document POD provides guidelines for use and the cleaning procedure for...

Страница 2: ...with White Port c 2 Red and Black 1 2m Cable Leads with Black Port d Disposable Electrodes e Calibrator f Measuring Tape 3 Contraindications This device cannot be used in participants with the followi...

Страница 3: ...ff device before replacing with 6 non rechargeable AA batteries only Duracell or Procell recommended 4 2 Connect Red and Black 1 2m Cable Leads with White Port to connector inputs at top of unit 4 3 A...

Страница 4: ...r participant s information NOTE Enter physical activity level using following table Level of Activity General Activities Very Low Movement Restricted Generally inactive Laying at ease sitting writing...

Страница 5: ...1 14 kph swimming 55 70 yds min 46 64 metres min 5 2 2 Screen will display Connect Electrodes 5 3 Participant Setup NOTE The following situations affect body water concentration Just prior during just...

Страница 6: ...een the medial and lateral malleoli must be minimum 3 cm apart 5 3 4 Shave electrode sites if necessary 5 3 5 Clean the sites with an alcohol swab 5 3 6 Allow sites to dry for 30 seconds before placin...

Страница 7: ...han previous value See example Frequency Khz 5 50 100 200 Impedance Values Ohms 588 505 475 454 8 Understanding Results 8 1 Body Composition only uses 50 kHz to predict Total Body Water TBW 8 2 Hydrat...

Страница 8: ...lt Estimated extra cellular water volume in litres ICW Estimated intra cellular water ICW lt Estimated intra cellular water volume TBW minus ECW BODY CELL MASS Total mass of cells in the body where ox...

Страница 9: ...required by certain types of BIA users for subject assessment REACTANCE 50 kHz PHASE ANGLE 50 kHz 10 Data Retrieval 10 1 Reading data from device 10 1 1 After successful measurements results are disp...

Страница 10: ...s 11 Safety Care and Maintenance 11 1 Safety 11 1 1 Do not connect QuadScan device to participants with active implanted medical devices i e cardiac pacemakers defibrillators or patients connected to...

Страница 11: ...vels Faulty Electrode Check Connection Cable leads have not been properly attached to the electrodes 1 Ensure participant is in correct position 2 Recheck the electrodes 3 Ensure leads are inserted pr...