Содержание V-shade AnchorFree V12

Страница 1: ... 1 40 ...

Страница 2: ...nstall Host Output 12 3 2 Disassemble Host Output 13 3 3 Stand 1 13 3 4 Install Handpiece and Change Handpiece 14 3 5 Install Earth Line 17 3 6 Fill Water 17 3 7 Drain Water 19 3 8 Place the Equipment 19 3 9 Start the Equipment 21 3 10 Main Interface 22 3 11 Error Noting 24 3 12 Wireless IC Card 26 3 13 Make Use of Wireless IC Card 28 3 14 Password 29 3 15 Treatment Procedure 30 3 16 Cryolipolysis...

Страница 3: ...34 4 5 Troubleshooting 35 Chapter 5 Clinical Guide 38 5 1 Working Mechanism 38 5 2 Applications 38 5 3 Customer Contraindications 38 5 4 Treatment Course 38 5 5 Customer s Feeling during Treatment 38 5 6 Treatment Skills 39 5 7 Treatment Part 39 5 8 Notes after Treatment 39 5 9 Suggestions for Beauty Salon 39 5 10 Questions and Answers 39 ...

Страница 4: ...system certification as the first in China With the efforts the vast majority of our products have been given medical CE certification we have established a strong R D center with nearly 30 people in R D team A number of our research results have filled the domestic blank and got patents among which 4 Results got invention patents 8 results got utility model patents Beijing Anchorfree actively pro...

Страница 5: ...tration Handpiece stand 3 Host output Handpiece stand 2 Screw Power switch Fuse Fill water Drain water Air exhaust Earth line connector Power supply Electricity breaker Standby handpiece connector Emergency switch Water level Working handpiece connector Stand 1 ...

Страница 6: ...6 40 1 1 Parts Host 1 168 handpiece 2 108 handpiece 2 Host output 1 Stand 1 Funnel 1 Earth line 1 Power line 1 Air exhaust with a hole 1 Filter User manual Warranty card ...

Страница 7: ...andpiece connects with right output connector right side of host screen is yellow and pilot lamp of handpiece shows yellow a Press key to adjust vacuum b After adjusting vacuum press key to start vacuum c You could also adjust vacuum on host screen d You have to start refrigeration on host screen e Press key to close refrigeration f After you close refrigeration and handpiece s temperature goes to...

Страница 8: ...168 handpiece connects with right output connector right side of host screen is yellow and pilot lamp of handpiece shows yellow a Press key to adjust vacuum b After adjusting vacuum press key to start vacuum c You could also adjust vacuum on host screen d You have to start refrigeration on host screen e Press key to close refrigeration f After you close refrigeration and handpiece s temperature go...

Страница 9: ...en 1 4 LCD screen Working mode Continuous Adjust scope 10minE90min Negative pressure scope 1 Absolute value 100kPa 20kPa 2 Relative value 0kPa 80kPa Vacuum levels 1E8 adjustable the higher the stronger Refrigeration levels 1E7 adjustable the higher the stronger Safety checking Real time on line Number of handpiece 4 handpiece 168 handpiece 2 ones 108 handpiece 2 ones Treatment area of tips 168mmx6...

Страница 10: ...nment request Avoid direct radiation of intense light Note 1 If the temperature in the room is too low or the equipment has not been used for any days you shall make the temperature of the whole system recover to 10 ºC or higher before starting it 2 You shall not use this equipment in humid environment Handpiece Treatment item 168 handpiece Stomach buttock waist leg 108 handpiece Back stomach thig...

Страница 11: ...onnel or the equipment 3 There is dangerous high voltage in the equipment So no one is allowed to disassemble any parts of the machine except the technical personnel appointed by our company 4 Good earthing is very important to operation safety This equipment has good earthing by yellow and green wires in power line Picture A 5 Power switch Picture B for turn on or turn off the equipment 6 Emergen...

Страница 12: ...12 40 Chapter 3 Operation 3 1 Install Host Output Picture 1 Picture 2 Picture 3 Picture 4 A C ...

Страница 13: ...efer to 3 3 Loosen C Picture 4 Pull out A Picture 1 Normally we don t disassemble the host output and we disassemble it while checking equipment or transportation 3 3 Stand 1 Picture 5 Installation Picture 5 Insert stand 1 into host output Normally the length is 50cm Fix the screw Adjustment Picture 5 Normally we don t adjust stand 1 If adjustment is needed first take out the handpiece pipe from D...

Страница 14: ... into D Disassemble Picture 5 First disassemble handpiece refer to 3 4 Put handpiece in a safe place Loosen screw Take out stand 1 3 4 Install Handpiece and Change Handpiece Handpiece connector 168 handpiece Handpiece connector 108 handpiece ...

Страница 15: ...15 40 Picture 6 Picture 7 Hndpiece connector Host output Unlock key Handpiece connector Host output ...

Страница 16: ... handpiece on stand 2 B in Picture 8 Place the standby handpiece on stand 3 C in Picture 8 Place the standby handpiece s connector like A in picture 8 Change handpiece Press the key Picture 7 at the same time pull out the handpiece connector Take out handpiece pipe from D Picture 5 Place the handpiece you just taken down in C Picture 8 and put its connector on B in picture 8 Insert working handpie...

Страница 17: ...arth line in the power outlet Installation Before turning on the equipment connect one side of the earth line with A following picture and connect the other side of the earth line to metal touching the earth like water pipe heater 3 6 Fill Water Use purified water or water without hydronium Strictly prohibit mineral water If you first use this equipment please take the following steps to fill wate...

Страница 18: ...p filling water 6 Press unlock key Picture 10 until you hear sound you pull out the funnel 7 Install the handpiece start the equipment and wait for 2 3 minutes 8 Turn off the equipment 9 Insert funnel into A Picture 10 until you hear the sound 10 Fill water 11 When water in the pipe gets the Water level sign Picture 11 stop filling water 12 Press unlock key Picture 10 until you hear sound you pull...

Страница 19: ...til you hear the sound 3 Put the funnel head in the container until there is no water flowing 4 Press unlock key Picture 10 until you hear sound you pull out the funnel 3 8 Place the Equipment Picture 12 禾 椦 Operator Customer Equipment Look at Picture 12 Take treatment on stomach for example Picture 12 shows the right place of customer equipment operator and the bed 1 Customer lie down the bed and...

Страница 20: ...tricity breaker Push A upward 5 Emergency switch shall not be pressed down 6 Press power switch 7 System enters initialization interface Picture 13 8 Then system enters password interface Picture 14 Original password is 123 9 Input password and system enters main interface Picture 15 Picture 13 Picture 14 A ...

Страница 21: ...21 40 3 10 Main Interface Picture 15 Picture 16 Picture 17 ...

Страница 22: ...n on vacuum switches on both sides of screen Then turn on refrigeration 2 After clinical treatment press left side refrigeration switch to turn off refrigeration H Key for handpiece working mode 1 Both handpiece Picture 15 2 Left handpiece L Picture 16 3 Right handpiece R Picture 17 I Right side refrigeration time J Right side handpiece type big handpiece is 168 small handpiece is 108 K Right side...

Страница 23: ...his part for long time to enter error introduction interfaces Picture 18 Picture 19 and Picture 20 Picture 18 Picture 19 Picture 20 3 11 Error Noting Error noting interface contains 3 interfaces Picture 18 Picture 19 and Picture 20 Press to change the 3 interfaces Press to go back to main interface Picture 15 EL1 Left handle isn t on line EL2 Left side has little water EL3 Left handle s NO 1 MOS p...

Страница 24: ...pe broke current ER5 Right handle s NO 1 MOS pipe short circuited ER6 Right handle s NO 2 MOS pipe short circuited ER7 Right handle s NO 1 temperature sensor has connection error broken isn t checked ER8 Right handle s NO 2 temperature sensor has connection error broken isn t checked ER9 The value deviation between NO 1 and NO 2 temperature sensors is more than 1 5 degrees ER10 Right handle s NO 3...

Страница 25: ...1 System enters Picture 22 2 Press or to set issued time 10 minutes for each press 100 minutes for each continual press Adjusted scope is 0 99900 minutes 3 Insert blank card to A in Picture 21 and system issues user card automatically Then system would note you Issue OK or Issue Error You can issue more user cards at the same time 4 After issuing user card press to enter Picture 22 ...

Страница 26: ...u Issue OK or Issue Error You can issue more user cards at the same time 4 After issuing user card press to enter Picture 22 Picture 23 Start IC Card Function Turn on the equipment insert issued user card to A in Picture 21 Equipment makes a sound and shows the left time in the user card Stop IC Card Function Turn on the equipment insert issued user card to A in Picture 21 Equipment makes a sound ...

Страница 27: ...e agent would issue a user card with 14 400 minutes again and give it to user If user wants to buy the equipment the agent could give key card to user after receiving payment Example 2 For beauty salon 1 The beauty salon takes charge by renting card treatment course and treatment times 2 10 times treatment course and 30 minutes time The beauty salon could take following measures to manage Turn on ...

Страница 28: ...e 24 3 In Picture 24 press the first and input original password Press the second and input new password Press the third and input new password again Press 4 Picture 25 shows password set successfully Picture 26 means original password is wrong Picture 27 means new password is wrong 5 If you don t want to change password press to go back ...

Страница 29: ... to 7 levels on both sides of screen 11 Let customer lie on the bed 12 Paste one piece of cloth on left side of customer s stomach and another one on right side of customer s stomach Picture 28 13 Press vacuum switches of both sides of screen Try vacuum level from low to high 14 Put one 168 handpiece on the middle of left cloth When skin is adsorbed totally you release your hand Picture 29 If skin...

Страница 30: ...matically Wait until handpiece s temperature goes to 8 degrees Centigrade pilot lamp on handpiece will flicker Lamp flicker means you can stop vacuum by hand Press key of handpiece connected to the orange screen Turn off vacuum switch of orange screen right side of screen Lightly take down handpiece of orange screen and put it on B in Picture 8 Press key of handpiece connected to the blue screen l...

Страница 31: ...fication Size Relative handpiece 1520 15cmX20cm Smaller handpiece HV08 3020 30cmX20cm Bigger handpiece HV07 The cryolipolysis cloth can only be used one time You can order the cryolipolysis cloth from our company Procedure P1 P2 P3 P4 P5 P6 P7 P8 P9 P10 P1 Find the small tearing of the package P2 Tear the package P3 Take out one piece of cryolipolysis cloth ...

Страница 32: ...e cover board opened up and no attendance Otherwise it can do harm to human being or the equipment itself 4 1 Add Water and Change Water in Time 1 Fill water once per month distilled water or water without ion Refer to 3 6 in Chapter 3 2 Change water once every 3 months Before changing water turn off the equipment Drain water Refer to 3 7 in Chapter 3 Fill water Refer to 3 6 in Chapter 3 4 2 Chang...

Страница 33: ...pressure sensor Take down the white cap Picture 35 If there is water in the air pipe dry the water Install the white cap again Picture 36 Screw down B Picture 34 Install the board A in Picture 31 Place the handpiece on the stands refer to Picture 8 Picture 31 Picture 32 Picture 33 Picture 34 A B B Filter B B ...

Страница 34: ...ter from handpiece Picture 37 2 Replace for a new filter Picture 37 4 4 Move the Equipment 1 On the floor hold the handle to move the equipment Picture 38 2 On the ladder two men lift the equipment by holding the board Picture 39 You must not lift the equipment by holding the handle Picture 38 Picture 39 Handle White cap White cap C ...

Страница 35: ...ction Touch screen has trouble Contact with our company Touch screen has no showing Touch screen has trouble Contact with our company There is no vacuum or less vacuum Air pump has trouble Contact with our company Handpiece leak air Contact with our company Form 2 Code Introduction Troubleshooting EL1 Left handle isn t on line Check the lines in handpiece EL2 Left side has little water Check the w...

Страница 36: ...efrigeration chip of left side EL17 Right side temperature of left handle is more than set value Change the refrigeration chip of right side ER1 Right handle isn t on line Check the lines in handpiece ER2 Right side has little water Check the water pipe ER3 Right handle s NO 1 MOS pipe broke current Change handpiece panel ER4 Right handle s NO 2 MOS pipe broke current Change handpiece panel ER5 Ri...

Страница 37: ...than set value Change the refrigeration chip of right side E1 Water temperature sensor has connection error broken isn t checkedChange water temperature sensor E2 Water temperature is more than 39 degrees Centigrade Standby and wait until water temperature is cool E3 Water temperature is more than 35 degrees Centigrade Standby and wait until water temperature is cool E4 Lower water level Fill wate...

Страница 38: ...dy Thus fat layer becomes thin 5 2 Applications Removal localized stubborn fat 5 3 Customer Contraindications The following customers shall not accept the treatment of this equipment 1 Patient with liver diseases 2 The skin to treated with has metal or some supports 3 Patient with skin diseases or any people with cancer 4 Other serious diseases like heart disease or diabetes 5 Women with menstrual...

Страница 39: ...tomer all the time 6 During treatment you have to turn off refrigeration firstly and then turn off vacuum 5 7 Treatment Part Stomach thigh leg waist Upper arm chin and other part with fat 5 8 Notes after Treatment 1 Use water with 15 E30 to wash treated skin in 2 hours after treatment 2 After washing treated skin deposit wet warm towel on the treated skin Note Don t use water more than 30 to wash ...

Страница 40: ...ated skin s fat is removed evidently 5 Customer How to take care of skin after treatment Beautician Use water with 15 30 to wash treated skin in 2 hours after treatment Then deposit wet warm towel on the treated skin for 10 20 minutes 6 Customer Why can t women with menstrual accept the treatment Beautician Normally women with menstrual are sensitive to pain and weak for cold 7 Customer Can anyone...