background image

 

 

     EN

 

 

 

   Assembly and operating instructions   

 

 

 

Nursing care beds 

 

 

 

 

 

 

 

domiflex 3 
 

Содержание Nursing domiflex 3

Страница 1: ...EN Assembly and operating instructions Nursing care beds domiflex 3...

Страница 2: ...for the provision of professional care The focus here is on the patient whose trust their health care providers seek to earn as they protect the patient s life and health With this health care product...

Страница 3: ...ch mode power supply 19 3 5 Box motor connections 20 3 6 Operating status display of the external SMPS switching power supply unit 20 3 7 Hand switch 21 3 8 Emergency lowering Back rest manual 23 Cons...

Страница 4: ...height extender 44 4 5 Tray 44 4 1 Stand up bar with attachment crossbeam 45 4 2 Mattresses 46 5 Cleaning care and disinfection 47 5 1 Cleaning and care 47 5 1 Disinfection 47 5 2 Avoiding hazards 48...

Страница 5: ...erm care in a medical environment in nursing care bed 4 In home care solely for use as a home care nursing care bed 5 Outpatient care 1 1 Intended use The nursing care bed is suitable for patients 146...

Страница 6: ...isinfection of the nursing care bed prior to reuse Please pay particular attention to all information about inspecting the bed Note The beds have no specific connectors for equipotential bonding Elect...

Страница 7: ...applicable national regulations on the obligations of the operator must be observed Point out the storage location of these instructions for use to the nursing staff in accord ance with the Medical D...

Страница 8: ...age hook and cannot fall to the ground The nursing care bed must only be operated by nursing staff or family members who have been trained in its use or in the presence of trained users When changing...

Страница 9: ...ly this nursing care bed has an ex pected lifespan of 10 years The actual lifespan will depend on the conditions and fre quency of use 1 7 Warranty For your warranty conditions for this nursing care b...

Страница 10: ...ion on a radio especially in direct proxim ity to the activated nursing care bed In these rare cases increase the distance between de vices use a different wall socket or temporarily switch off the de...

Страница 11: ...rrent consumption 5 Duty cycle 6 Antriebsschutzart 7 Own weight of the bed 8 Manufacturer 9 Maximum patient weight Safe operating weight 1 11 Nameplate with reinforcing struts Each reinforcement strut...

Страница 12: ...be disposed of in household waste Type B medical applied part Use only in dry areas Protection class II double insulation protective insulation IPX4 Electrical equipment protected from splashing wate...

Страница 13: ...cial spring support systems and is divided into four functional zones Back rest fixed seat thigh rest and lower leg rest The outer edge of the mattress support frame is constructed from tubular steel...

Страница 14: ...ows for easy adjustment of the side rails 2 1 Adjusting the continuous side rail The release button for adjusting the continuous side rails is lo cated just above the upper side rail in the sliding tr...

Страница 15: ...ermissible range Ensure that the side rails click securely into position Before placing the side rails and before each new use check for damage on all mechanical parts on the bed frame and the side ra...

Страница 16: ...ription A Gap between headboard and side rail B Side rail height 1 C Side rail height 2 D Side rail length 1 E Distance between elements of the side rail system F Distance between separated side rails...

Страница 17: ...s protection meets the standards of protection code IPX4 The maximum duty cycle is indicated on the nursing care bed nameplate For example a 10 duty cycle 2 min ON 18 min OFF means that each electri c...

Страница 18: ...a result of improper cleaning Such damage can result in control unit malfunction which can in turn lead to unintended move ments of individual bed components raising the risk of injury to patients an...

Страница 19: ...r supply unit and an emergency low ering mechanism The SMPS switch mode power supply is an electronic transformer that warms up only minimally under load and has an integrated electronic power monitor...

Страница 20: ...The SMPS switched mode power supply has an LED that can indicate the following oper ating states LED On Ready for operation LED Off Discharged not connected LED Flashing Error thermal overload or sho...

Страница 21: ...hanger can be used to hang the hand switch on the nursing care bed e g during cleaning or when providing patient care To keep the hand switch from getting in the way simply clip it anywhere on the be...

Страница 22: ...ut all electrical functions place the key in the lock on the back of the hand switch and turn the key to turn the locking function on or off Locking mechanism 1 Standard Locking mechanism 2 option wit...

Страница 23: ...k rest and can be pushed downward Once the second person has left the danger zone the first person can carefully lower the back rest Be sure to keep a firm grip on the back rest until it has been full...

Страница 24: ...vel dB A 65 65 65 Adjustment angle Back rest 70 70 70 Thieg rest 42 42 42 Lower leg rest 16 16 16 Trendelenburg position 15 15 15 Weights Total incl wooden side rails kg 73 9 75 8 77 8 79 7 80 8 82 7...

Страница 25: ...L 103 x 90 7 x 210 7 113 x 90 7 x 210 7 Safe operating weight kg 220 220 Max patient weight kg 185 185 Height adjustment range cm 35 80 35 80 Back rest length cm 66 66 Mattress compensation length cm...

Страница 26: ...x 3 model is not intended for use in hospitals can in some circumstances be combined for medical purposes when needed with other electrical medical devices e g aspirators ultrasonic nebulisers feed in...

Страница 27: ...EN 27 3 2 Assembly and installation video Mounting video domiflex 3 Extension video bed extension Extension video reinforcing stuts Transport...

Страница 28: ...upport frame which has no mounting bracket for the trapeze bar Attach the mattress support frame section to the end piece by sliding the mattress support frame over the mounting tabs The end piece can...

Страница 29: ...ion Locate the head section of the mattress support frame Put the lying surface on the head side onto the lying surface on the foot side the tool free connection must snap into place The lying surface...

Страница 30: ...cable connectors should be fac ing toward the centre of the bed Insert the cover plates sideways into the slide locks on the box motor Locate the second end piece Push the end piece into the head sect...

Страница 31: ...d plugs indicating which is the top rail and which is the bottom rail The lower rail is also taller than the upper rail Pull all side rails upward Then push the end piece into the second detent positi...

Страница 32: ...the mounting bracket Ensure that it is locked in place with the groove Test the operation of the nursing care bed includ ing adjusting the position and height of the mattress support frame Your domif...

Страница 33: ...cing struts take the reinforcing struts out of the pack aging Hang the reinforcing struts on the bed frame and attach them with the screws provided Tighten the screws fully Make absolutely sure that t...

Страница 34: ...on nector pieces each with a tube clip to make a connection between the end pieces Arrange the end pieces to be slightly off set from one another Push both end pieces in Place the tube clip as shown i...

Страница 35: ...mantle the box motor by loosening the slides and then putting it aside Next insert the foot section of the mat tress support frame from above The mat tress brackets should face upward and outward Then...

Страница 36: ...EN 36 Visually compare the assembly with the figure at right The side rail system can now be inserted Next insert the side rail holder into the head section of the mattress support frame...

Страница 37: ...g free flat steel The erector should face inwards Be careful not to damage the tool free connection Now take a springwood slat out of the moulding on one side and place the box motor between the lying...

Страница 38: ...l the end piece out of the lying surface to the first detent The side guards are also released by this and can be put aside Then remove the end piece completely Insert the frame extensions and lock th...

Страница 39: ...der so that it is level with the foot bar Once you have finished this step your as sembly of the bed extension is complete 3 7 Disassembly Before starting to disassemble the bed unplug it from the wal...

Страница 40: ...be damaged by moving parts of the nursing care bed then put the plug back into the socket 3 9 Transport storage and operating conditions Transport and storage Operation Temperature 0 C to 40 C 10 C t...

Страница 41: ...always be entrusted to qualified professionals Malfunction Possible causes Suggestion Drive system cannot be oper ated via hand switch Power cable unplugged Plug in the power cable No power to wall so...

Страница 42: ...r patients with less common body types optimal comfort can only be achieved with special design variants With its options for beds with non standard dimensions Her mann Bock GmbH makes it possible to...

Страница 43: ...rmann Bock GmbH NOTE Do not swing the trapeze bar outside of the bed perimeter Under normal use conditions the triangle grip has a lifespan of approximately 5 years If a trapeze bar with triangle grip...

Страница 44: ...ender weighs 1 0 kg Includes Side rail height extender fully assembled Open plastic closure place height extender on top of rail position in the centre of the rail and close the closure Be sure that t...

Страница 45: ...kg Includes Stand up bar with attachment crossbeam and installation hardware Hold the attachment crossbeam against the mattress support frame from below and mark the drill holes on the frame with a p...

Страница 46: ...rames with spring systems For thicker mattresses an additional rail accessory side rail height extender available as a product accessory must be used as well The mattress thickness must not be less th...

Страница 47: ...damp cloth with no added cleansers is generally sufficient When caring for plastic surfaces use products that are specifically intended for use on plastic Drive system To prevent moisture from enterin...

Страница 48: ...ow it dry if necessary If you suspect that moisture may have got into the electrical components imme diately pull out the wall plug and or do not plug it back into mains power under any circumstances...

Страница 49: ...V 8kV 15kV Contact discharge 8 kV Air discharge 2 kV 4kV 8kV 15kV Floors should be made of wood or concrete or be covered with ceramic tiles If the floor is covered with syn thetic material the relati...

Страница 50: ...jects and people The field strength of fixed transmitters e g base stations for mobile phones and land mobile radios amateur radio stations AM and FM radio broadcasters and TV broadcasters cannot be t...

Страница 51: ...ntation may only be done by or under the supervi sion of qualified professionals such as qualified electricians or people with technical train ing in electrical systems who are familiar with the appli...

Страница 52: ...ds and plugs free of breaks kinks or pinched areas abrasions porous areas and exposed wires 10 Strain relief device firmly screwed into place and working properly 11 Safe correct cable management 12 M...

Страница 53: ...han 31 8 cm 34 Side rails slide easily in tracks and lock securely into place For dino bed only Gates slide easily along aluminium profiles Gates lock securely into locking mechanism 35 Side rails sec...

Страница 54: ...d other forms of service can only be provided by our authorised personnel and sales partners Otherwise all warranty claims will be void For a list of our current sales partners see www bock net kontak...