EO2K, EO4K and EO6K Diffusion Pumps
IN
STALLATION
PAGE
27
Issue G
27
Aug 06
Figure 14 – Cooling-water connections
3.8
Leak-test the system
The EO pumps are stringently leak-tested before despatch. After you have installed the pump, leak-test
the system before you fill the pump with pump fluid, and seal any leaks found; outgassing from the fluid
may give false test results. The pump fluid is viscous at ambient temperatures and may block small leaks.
The level of leak tightness required will depend on the application of your system. You must leak-test the
system to ensure the integrity of your vacuum system and its vacuum seals.
A
EO2K
B
EO4K
C
EO6K
1.
Cooling-water inlet
2.
Cooling-water outlet
Содержание EO2K
Страница 1: ...B304 09 880 Issue G Original Instruction Manual EO2K EO4K and EO6K Diffusion Pumps ...
Страница 2: ...This page intentionally blank ...
Страница 6: ...EO2K EO4K and EO6K Diffusion Pumps PAGE iv Issue G iv Aug 06 This page intentionally blank ...
Страница 14: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 8 Issue G 8 Aug 06 This page intentionally blank ...
Страница 20: ...EO2K EO4K and EO6K Diffusion Pumps TECHNICAL DATA PAGE 14 Issue G 14 Aug 06 Figure 10 EO6K dimensions mm ...
Страница 22: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 16 Issue G 16 Aug 06 This page intentionally blank ...
Страница 36: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 30 Issue G 30 Aug 06 This page intentionally blank ...
Страница 52: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 46 Issue G 46 Aug 06 This page intentionally blank ...
Страница 54: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 48 Issue G 48 Aug 06 This page intentionally blank ...
Страница 58: ...EO2K EO4K and EO6K Diffusion Pumps PAGE 52 Issue G 52 Aug 06 This page intentionally blank ...
Страница 61: ...This page intentionally blank ...