10
Instruction: The default setting will display after the first booting display, and the contents and measurements set by
the user prior to the previous shutdown will display after the next booting.
3.3 Stylus position
In main screen, press Return button to check the stylus position of the pickup. In general, for measuring in the
plane, it is not necessary to adjust the position of the stylus. After measuring, the error of results is within the range
of the instrument.
The Stylus Position screen is used for the column, to adjust the height of the instrument up/down as required, in
order to determine whether the pickup has contacted the measured surface. The measuring may begin as the
point is near the zero point.
3.4 Calibration
Before the measuring, calibration of the instrument using Calibration specimen is usually required. This instrument
is accompanied with a calibration specimen. Before the measuring, test the calibration specimen using the
instrument. Normally, as the difference between the measurement and the value of the specimen is within the
range of the instrument, the measurement is valid. It is unnecessary to adjust. The tester can be used for direct
measuring.
If the difference between the measurement and the value of the calibration specimen exceeds the range of the
instrument, or the user requires higher measurement accuracy, you can use the calibration function to correct the
measurements to improve the accuracy.
Calibration procedure:
In main screen, press and hold Set button to enter the main menu. Press Up/Down to select “Calibration”. Press
Return to enter Display Calibration Setting screen.
Press Up/Down to change the row. Select the desired range to calibrate. Press Left/Right to adjust the setting,
then press Return to confirm the change.
Estimate the value for correction depending on the error of the measurements. Input the measurement range and