Bms Bulut Makina SR210 Скачать руководство пользователя страница 11

 

11 

 

calibration value can only calibrate the selected measurement range. Under this situation, press measurement range 
key (Range) can calibrate the indicating values respectively of different measurement ranges. Press the measurement 
range/indicating  value  calibration  key  again  after  setup  the  calibration  value  to  drop  out  indicating  value  calibration 
situation, the "CAL" symbol would disappear, and could take normal measurement. 

Means the calibration is carried according to 3% of actual deviation.

 

Means the calibration is carried according to -5% of actual deviation.

 

Instruction:    1. When testing model of the instrument by correct measurement method, if the actual measurement value 

exceeds ±10% of the model calibration value, the indicating value calibration function can calibrate according to 
the percentage of actual deviation, and the calibration range is no bigger than ±20%. 
2. Usually all instruments would test strictly before leave factory, the indicating value deviation is much less than 
±10%. According to above information, please do not use the indicating value calibration function frequently. 
3. If the indicating value calibration value is set as "00", all setting calibration would remove, and the instrument 
would come back to factory settings. 

3  Optional accessories and the application   

3.1  Height support and probe holder 

If  the  test  surface  of  the  work  piece  is  smaller  than  the  under-surface  of  the  instrument,  please  use  the  instrument 
accessories such as probe holder and height support as auxiliary supporting so as to finish the measurement (As shown 
in the below figure) . 

 

The application of height support and probe holder 

Note:    1.  The  bearing  surface  should  not  be  smaller  than  the  drive  route  of  this  measurement  so  as  to  avoid  the 

situation that probe falls away from the work piece when measuring, and stands up to the work piece when get 
back. 
2. The lock of height support should be reliable. 
3. Please do not use height support to adjust stylus position when measuring. Please adjust the height of variable 
stand  bar  to  the  required  height  before  measurement.  Usually  measure  through  calipers  can  satisfy  the 
requirement. 

-

+

0

CAL

RC

m

Ra

±80

±40

±20

2.5

0.8

0.25

Ra

m

RC

CAL

0

+

-

0.25

0.8

2.5

±20

±40

±80

Содержание SR210

Страница 1: ...R210 BMS Bulut Makina Sanayi ve Ticaret Ltd Şti Kocaeli KOBİ OSB Köseler Mahallesi 6 Cadde No 20 2 Dilovası KOCAELİ TURKEY Phone 90 262 502 97 73 76 90 262 503 06 51 Web www bulutmak com e mail bms bulutmak com OPERATION MANUAL ...

Страница 2: ...4 SYSTEM SETTING 10 2 4 1 Unit system 10 2 4 2 LCD Back light 10 2 5 FUNCTION SELECTION 10 2 5 1 Stylus position 10 2 5 2 Calibration of indicating value 10 3 OPTIONAL ACCESSORIES AND THE APPLICATION 11 3 1 HEIGHT SUPPORT AND PROBE HOLDER 11 3 2 HEIGHT STAND 12 3 3 EXTENSION ROD 12 3 4 PROBE 12 3 4 1 General purpose probe 12 3 4 2 Curved surface probe 12 3 4 3 Small bore probe 13 3 4 4 Super small...

Страница 3: ...RS DEFINITIONS 17 7 2 1 Arithmetic Mean Deviation of Profile Ra 17 7 2 2 Root mean square Deviation of Profile Rq 17 7 2 3 Maximum height of the profile Rz 17 7 2 4 Total Peak to valley Height Rt 18 8 SAMPLING LENGTH SELECTION RECOMMENDATION FORM 18 ...

Страница 4: ... time long continuous working hour Mechatronics design with small dimension and light weight easy to carry Auto power off function can save battery Optional measurement platform small bore probe curved surface probe extension rod and other accessories 1 1 Measurement Principle When measuring the work piece surface roughness please put the probe on the test surface of work piece The internal actuat...

Страница 5: ... Description Picture Description Picture Main unit Height support General purpose probe Calibrate specimen Probe holder Calibrate specimen holder Screw driver Charger 1 3 Layout of the instrument 1 3 1 Main unit 1 3 2 Probe ...

Страница 6: ... lr Press one time change one sampling length value Measurement range change Indicating value calibration Range Press one time change one measurement range long press enter into indicating value calibration process 1 4 Basic connection method 1 4 1 Assembling and dis assembling of the probe As shown in the figure hold the connection part of top guide tray and probe main body stylus is downward plu...

Страница 7: ...ll charge 3 The battery switch is off when the instrument leave factory 4 If the instrument work in properly and turn off turn on the machine cannot help to solve the problem then please turn off the battery switch of the instrument backside and restart after 10seconds 2 Measurement Operation 2 1 Preparation before measurement a Please check if the battery voltage is regular or not after start up ...

Страница 8: ...ement situation Probe is sliding Probe gets back Note If the instrument is turned off by accidental touch of power key during measurement situation the instrument probe would turn back to initial position after restarting up Please do not carry out any operation to instrument probe during this time until the probe back to its last position 2 3 Measurement Conditions Modification 2 3 1 Sampling len...

Страница 9: ...s of RC PC RC Gauss and D P until it select the required filter according to the filter indicating symbol of the LCD top place Select RC Select PC RC Select GAUSS Select D P 2 3 4 Parameter After press the parameter key Ra the instrument can change periodically on the 4 parameters of Ra Rz Rq and Rt until it select the required parameter according the display of the LCD Select Rz Select Ra Ra m RC...

Страница 10: ...and instrument try to make sure that the stylus position is near to 0 so as to obtain best measurement result All measurements are valid unless the stylus position exceed the measurement range During the measurement process please make sure that the probe main body is parallel to the test surface visual inspection is enough At this time all measurements are valid if the stylus position is within t...

Страница 11: ...ng to above information please do not use the indicating value calibration function frequently 3 If the indicating value calibration value is set as 00 all setting calibration would remove and the instrument would come back to factory settings 3 Optional accessories and the application 3 1 Height support and probe holder If the test surface of the work piece is smaller than the under surface of th...

Страница 12: ...ht it can precisely adjust the stylus position and the measurement would be more stable Please use upright if the test surface Ra value is small Please use upright if use professional probe such as small bore probe deep notch probe and curved surface probe Height stand 3 3 Extension rod Use extension rod can increase the depth which the probe enters into the work piece inside The length of extensi...

Страница 13: ...13 3 4 3 Small bore probe 3 4 4 Super small bore probe 4 Technical parameters ...

Страница 14: ...tion instrument after scratch 5 1 Fault treatment If there are any fault for the instrument please process according to measures which stated on the following chapter of fault message please send it back to factory for repair if it still could not work properly Please do not take down and repair by yourselves Please attach the warranty card and standard model of the instrument together with the in...

Страница 15: ...urt line calculation from original profile 7 1 2 Filter RC It is the traditional two level RC filter and conform to old standard Considering that some users are still using it so we keep it as transition The input and output signal of this instrument has phase difference PC RC It is the filter which has taken digital phase correction on the basis of RC filter It s amplitude transmission feature is...

Страница 16: ...gures when using RC Filter PC RC Filter GAUSS Filter Traversing length Approach Travel 0 5mm Pre travel l Assessment length l n Origin End Pre travel l Assessment length l n Post travel l Traversing length Origin End Approach travel 0 5mm l λc l λc Pre travel lr 2 Post travel l 2 Traversing length Origin End Approach travel 0 5mm Assessment length l n l λc ...

Страница 17: ...a 1 1 y i y Ra 0 X l 7 2 2 Root mean square Deviation of Profile Rq The root mean square value of the vertical coordinate value within one sampling length 2 1 1 2 1 n i i y n Rq 7 2 3 Maximum height of the profile Rz The sum height between maximum profile peak height Rp and maximum profile valley depth Rv within one sampling length Approach travel 0 5mm Assessment length l n Traversing length Orig...

Страница 18: ...th selection recommendation form Ra µm Rz µm Sampling length mm 5 10 20 40 2 5 2 5 5 10 20 1 25 2 5 6 3 10 0 8 0 63 1 25 3 2 6 3 0 32 0 63 1 6 3 2 0 25 0 32 1 25 1 6 0 25 0 20 0 25 1 0 1 25 0 16 0 20 0 8 1 0 0 125 0 16 0 63 0 8 0 1 0 125 0 5 0 63 0 08 0 1 0 4 0 5 0 063 0 08 0 32 0 4 0 05 0 063 0 25 0 32 0 04 0 05 0 2 0 25 0 032 0 04 0 16 0 2 0 025 0 032 0 125 0 16 0 02 0 025 0 1 0 125 ...

Отзывы: