Membrane:

 ME/TH/CO test strip:  Methamphetamine, THC and Cocaine-protein 

conjugates are coated onto specific region on the membrane known as the “Test 
Region”.   

AM/OP/PC or AM/OP/BZ test strip:  Amphetamine, Morphine, Phencyclidine or 
Amphetamine, Morphine, Benzodiazepine protein conjugates are coated onto the 
test region of the membrane.

 

Oratect

®

 III Oral Fluid Drug Screen Device 

Colloidal Gold Conjugate Pad:

  The colloidal gold conjugate pad for the ME/TH/CO 

test strip contains anti-methamphetamine, anti-THC and anti-cocaine antibody 
colloidal gold conjugates coated onto a fibrous pad.  The colloidal gold conjugate pad 
for the AM/OP/PC or AM/OP/BZ test strip contains anti-amphetamine, anti-morphine, 
anti-phencyclidine or anti-amphetamine, anti-morphine, anti-benzodiazepine antibody 
colloidal gold conjugates.  

ME/TH/CO/AM/OP/PC or BZ 

Catalog No. HM11 and HM12

 

 
Intended Use 

The Oratect

®

 III Oral Fluid Drug Screen Device is a one-step lateral flow 

immunoassay device for the qualitative detection of methamphetamine, MDMA, THC, 
cocaine, amphetamine, opiates, phencyclidine or benzodiazepines in human oral 
fluid. The Oratect

®

 III Test detects these drugs at the following cut-off concentrations: 

Collection Pad:

  The collection pad consists of an absorbent material. 

Materials Provided

 

Each Oratect

®

 III Oral Fluid Drug Screen Device kit contains: 

ME 
TH 
CO 
AM 
OP 
PC 
BZ 

d-Methamphetamine/MDMA 
Delta-9-Tetrahydrocannabinol 
Cocaine 
d-Amphetamine 
Morphine 

Phencyclidine 

    Diazepam 

25 ng/ml   
40 ng/ml 
20 ng/ml 
25 ng/ml 
10 ng/ml 

4 ng/ml 

           5 ng/ml

1.  1 Package Insert. 
2.  1 Reference Guide. 
3.  25 test devices.  Each device consists of a plastic holder and a detachable 

cap.  The devices are packaged individually in a foil pouch with a desiccant. 

4.  1 plastic vial containing buffer for confirmation test. 

 

Materials Required but Not Provided 

• 

Timing device  

 

 

The test is intended to be administered by a trained professional.  It should not be 
used without supervision.  This product is intended for forensic use only and is not 
for use in diagnostic procedures. 

Warnings and Precautions

 

• 

The Oratect

® 

III Oral Fluid Drug Screen Device is intended for forensic use only 

and is not for use in diagnostic procedures. 

The Oratect

®

 III Oral Fluid Drug Screen Device provides only preliminary drug 

test results. For a quantitative result or for a confirmation of a presumptive 
positive result obtained by the Oratect

®

 III Oral Fluid Drug Screen Device, a 

more specific alternative method such as GC/MS or LC/MS must be used.  

• 

The test device should remain in its original sealed pouch until ready for use.  

• 

Discard the test device if package is ripped or torn. 

• 

Do not use the test device beyond the expiration date indicated on the kit. 

• 

Handle all oral specimens as potentially infectious. Proper handling and 
disposal methods should be established. 

 

Summary and Explanation

 

Illegal drug consumption contributes to many accidents, injuries and medical 
conditions. Screening individuals for drugs of abuse is an important method in 
identifying those who may cause harm to themselves and to others. 

Product Storage 

The Oratect

® 

III Oral Fluid Drug Screen Device pouch should be stored at room 

temperature (15°-30°C). Do not open pouch until ready to perform the assay. 

Oratect

®

 III Oral Fluid Drug Screen Device is developed to detect active drugs-of-

abuse present in saliva. Studies on methamphetamine, MDMA, cocaine, opiate, 
amphetamine, phencyclidine, benzodiazepine and cannabinoid show that all of these 
drugs are detectable in oral fluid. Oratect

®

 III Oral Fluid Drug Screen Device is 

designed to integrate oral fluid collection and  lateral flow immunoassay screen 
testing for drugs-of-abuse in one single device. 

Specimen Collection and Handling 

IMPORTANT:  At least 10 minutes prior to administering the test, instruct the donor 
not to eat, drink, smoke or chew tobacco products. 

Test Procedure 

1. 

Remove the test device from the sealed pouch. 

2. 

Carefully remove the blue cap by holding the sides and pull gently.  This will 
expose the collection pad. 

Test Principle

 

The Oratect

®

 III Oral Fluid Drug Screen Device is based on a competitive 

immunoassay procedure in which drug derivatives immobilized on the membrane 
compete with the drug(s) which may be present in oral fluid for limited antibody 
binding sites on the colored colloidal gold antibody conjugate. During testing, oral 
fluid is collected at the collection pad and migrates across the membrane. If no drug 
is present in the oral fluid, the colored colloidal gold antibody conjugate will bind to 
the drug derivatives on the membrane to form visible bands at specific test regions. 
Therefore, the 

presence of a purple-red band

 at a specific test region indicates a 

negative

 

result

. If any drug(s) is (are) present in the oral fluid, it competes with the 

immobilized drug conjugate for limited antibody binding sites of the colored colloidal 
gold conjugate. When a sufficient amount of drug is present, the drug will saturate 
the antibodies, and the colored colloidal gold conjugate cannot bind to the drug 
derivative on the membrane. 

Therefore, the absence of a purple-red band at the 

test region indicates a presumptive positive result for that particular test. 

3. 

Ensure that the blue line is present in each test window. 

4. 

The oral fluid collection process must be observed. Instruct the donor to hold 
the top portion of the device (above the test windows).  

5. 

When placing device into the mouth, 

keep head level

a. 

Open mouth and rub the collection pad inside mouth against one cheek 
gently in a circular motion several (approximately 15-20) times. 

(Fig. b)

 

b. 

Still keeping head level, gently rub the collection pad against the opposite 
cheek in a circular motion (approximately 15-20) several times.  

(Fig. b)

 

 

 

Fig. b Gently rub the collection pad against each cheek several (approximately 15-20) times. 

c. 

Rub the collection pad on top of the tongue several times and then 
underneath the tongue several (approximately 15-20) times. 

(Fig c. and 

Fig d.).  Do not chew, suck, bite or bend the collection pad. 

 

 

Fig. c   Gently rub the collection pad on 
top of the tongue several (approximately 
15-20) times. 

Fig. d   Gently rub the collection pad 
underneath the tongue several 
(approximately 15-20) times. 

 

Fig. a    Detail regions of 

Oratect

®

 III Oral Fluid Drug Screen Device 

The presence of a blue line in each window indicates that the device is unused. The 
flow of the blue lines indicates that a sufficient amount of oral fluid has been 
collected.  A control band at the control region (C) indicates the test has performed 
properly. This control band should always appear regardless of the presence of drug 
or metabolite.  

Reagents

 

The Oratect

®

 III Oral Fluid Drug Screen Device contains two membrane strips and a 

collection pad. Each strip consists of a membrane, a colloidal gold conjugate pad, a 
sample pad and an absorbent pad. 