M1 Mini User Manual
20 / 32
When there is a new version of LightTrip App or device firmware, there will be a prompt when
opening LightTrip App. According to the prompt, the software and firmware can be upgraded.
CAUTIONS!
• Please maintain Bluetooth connection between APP and main device during firmware
upgrade.
• In order to ensure the best performance, it is recommended to keep the latest version of
LightTrip App and device firmware.
18. Cleaning and Maintenance
WARNINGS!
• Regular cleaning of the device and its accessories is very important for the prevention of
respiratory infections.
• To avoid electric shock, always unplug the device before cleaning.
• Use washing liquid that is nontoxic to humans and does not cause allergies in humans.
• Follow the manufacturer's instructions on cleaning the mask and tube and on determining
the frequency of cleaning.
• Do not open or modify the device. There are no user serviceable parts inside. Repairs and
servicing should only be performed by an authorized service agent.
• In order to prevent cross-infection of patients or contamination of equipment, BSF (Breathing
System Filter) that meet ISO 23328-1:2003 and ISO 23328-2:2002 standards and have
medical device registration certificates can be used.
(1) Different patients need to replace a new BSF before using this equipment.
(2) When using the BSF, please install and operate it according to the instructions of the BSF,
and pay attention to adjust the output pressure setting of the device according to resistance of
the BSF to ensure delivering normal treatment pressure.
(3) Atomization or humidification will increase the resistance of the BSF. The operator must
often monitor the resistance increase and blockage of the BSF to ensure delivering normal
treatment pressure.
• If you use ozone or other cleaning and disinfection methods that are not recommended by
BMC, BMC will not be able to verify the safety or performance of the equipment.
CAUTIONS!
• Overheating of the materials could lead to early fatigue of these materials.
• Do not use solutions containing chlorinated lime, chlorine, or aromatic to clean the device