Guidance and manufacturer's declaration - electromagnetic immunity

 

The RESmart

®

 is intended for use in the electromagnetic environment specified 

below. The user of the RESmart

®

 should make sure that it is used in such an 

environment.  

Immunity 
test

 

IEC 60601 
test level

 

Compliance 
level

 

Electromagnetic 
environment - guidance 

 

Electrostatic 
discharge 
(ESD)  
 
IEC 61000-4-2 

±6 kV contact 
 
 
 
±8 kV air   

±6 kV 
contact  
 
 
±8 kV air   

Floor should be wood, concrete 
or ceramic tile. If floors are 
covered with synthetic material, 
the relative humidity should be 
at least 30%.   

Electrical fast 
transient/burst 
 
IEC 61000-4-4 

±2 kV for   
power supply 
lines 
 
±1 kV for   
input/output 
lines 

±2 kV for   
power supply 
lines 
 
±1 kV for   
input/output 
lines 

Mains power quality should be 
that of a typical home or 
hospital.  

Surge  
 
IEC 61000-4-5 

±1 kV   
differential 
mode 
 
±2 kV   
common 
mode 

±1 kV   
differential 
mode 
 
±2 kV   
common 
mode 

Mains power quality should be 
that of a typical home or 
hospital.  

Voltage dips, 
short 

interruptions 
and voltage 
variations on 
power supply 
input lines   
 
IEC 
61000-4-11  

<5 % 

U

T

   

(>95% dip in 

U

T

)  

for 0.5 cycle 
 
40% 

U

T

   

(60% dip in 

U

T

)  

for 5 cycles   
 
70% 

U

T

   

(30% dip in 

U

T

)  

for 25 cycles 

 
<5% 

U

T

   

(>95% dip in 

U

T

)  

for 5 s   

<5% 

U

T

   

(>95 % dip in 

U

T

)  

for 0.5 cycle   
 
40% 

U

T

   

(60% dip in 

U

T

)  

for 5 cycles   
 
70% 

U

T

   

(30% dip in 

U

T

)  

for 25 cycles   
 
<5% 

U

T

   

(>95% dip in 

U

T

)  

for 5 s   

Mains power quality should be 
that of a typical commercial or 

hospital environment. If the 
user of the RESmart

®

 requires 

continued operation during 
power mains interruptions, it is 
recommended that the 
RESmart

®

 be powered from an 

uninterruptible power supply or 
from a battery.   

Содержание INH2

Страница 1: ...English InH2 Heated Humidifier User Manual 0123...

Страница 2: ...er is only used for single patient and must not be re used on another person This is to avoid the risk of cross infection The InH2 Heated Humidifier is not intended for use with a patient whose upper...

Страница 3: ...only accessories recommended by BMC Never operate the humidifier if any of the parts are damaged if it is not working properly or if the humidifier has been dropped or mishandled Do not use the humidi...

Страница 4: ...indicates the maximum water level for safe operation Outlet Port Connect the flexible tubing coaxial 22 mm here Water Chamber The removable water chamber holds the water for humidification It has a s...

Страница 5: ...ector Plug this connector into the power outlet on the BMC RESmart Device Set Up 1 Connect the Humidifier to the RESmart device Make sure they are connected completely 2 Place the RESmart device and h...

Страница 6: ...ifier will work automatically The yellow indicator light on the humidifier will turn on Press the humidifier button can turn off or restart the humidifier 5 The ideal humidity setting depends on room...

Страница 7: ...mes with clean water and allow to air dry 4 Fill the water chamber and close the silica gel cover Inspect the water chamber for any leak or damage Replace the water chamber if any damage is present Hu...

Страница 8: ...condensing Atmospheric Pressure 860 to 1060 hPa 500 to 1060 hPa Power Requirements 24V DC 1 0A max Type of Protection Against Electric Shock Class II Equipment Degree of Protection Against Electric Sh...

Страница 9: ...the System When traveling the optional CPAP carrying case is for carry on luggage only The carrying case will not protect the humidifier if it is put through checked baggage Security Stations For eas...

Страница 10: ...missions CISPR 11 Group 1 Class B The RESmart uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic eq...

Страница 11: ...for power supply lines 1 kV for input output lines Mains power quality should be that of a typical home or hospital Surge IEC 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2...

Страница 12: ...re than is indicated in the device specification it may be necessary to position the RESmart further from sources of power frequency magnetic fields The power frequency magnetic field should be measur...

Страница 13: ...nications equipment should be used no closer to any part of the RESmart including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transm...

Страница 14: ...adio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey sh...

Страница 15: ...ive material or part BMC Medical Co Ltd will pay customary freight charges from BMC Medical Co Ltd to the dealer location only This warranty does not cover damage caused by accident misuse abuse alter...

Отзывы: