BMB MEDICAL – CLAVIA 5 / 1VCLVE092020
86
2.15.10
IV POLE
The removable IV POLE is placed in one of the 2 holders placed on either side of the barrier
(LSA/ LSA SURGERY/FSP version) or the armrest (FCD version).
In each holder, three locations are available in order to adapt the position of the IV POLE to the
positioning of the chair.
To deploy the IV pole:
3-
Turn the thumb wheel
counter-clockwise and extend the
telescopic part of the IV pole
to the desired height.
4-
To lock the telescopic part, turn the thumb wheel
clockwise.
Ensure that the IV pole is securely locked.
Safe working load: 6 kg
.
This indicative value corresponds to safe product use, but is not representative of
normal use for which the load is significantly lower.
When moving the chair fitted with the IV pole, ensure that the latter does not strike
any obstacles (e.g. door frame, lighting, etc.) or individuals.