BMB MEDICAL – CLAVIA 5 / 1VCLVE092020
69
“Warning: Do not short-circuit the batteries and do not leave the system unused and without
charging for extensive periods of time as this could cause self-discharge, leading to medium-
term damage. This will also have an environmental impact. »
Battery characteristics
Battery safety symbols
Battery type:
Lithium-Ion
IPX6
Protection rating
Capacity:
2.25 Ah
CE marking
Input supply voltage:
29~45 VDC
Indoor use
Input supply voltage:
25.9 VDC, 10 A
Electronic waste
2.4
PRODUCT
POWER
SUPPLY
Power grid connection conditions and electromagnetic compatibility:
This medical device’s power cord may represent an obstacle that could cause a fall. Be
mindful of its presence when moving around the appliance, when the latter is connected to
the mains.
Position the product such that it can be easily unplugged.
Unplug the product before moving it.
Do not leave the cables in contact with the floor or the casters,
Do not knot the cables,
Periodically check the condition of the cables.
Mains connection:
100 ~ 240 V depending on the country.
Frequency:
50 /60 Hz
Rated current:
3.3 A
Protection rating / degree of electric shock
protection:
Class
2
appliance
Power input:
525 W
Intermittent operating mode:
5 %, i.e. 1 minutes of continuous operation
followed by 19 minutes of rest.
The mains plug is the electrical grid insulation device.
Work on electrical parts should only be performed by qualified and authorized staff.
Electromagnetic interference caused by the chair cannot be excluded. In this case, the chair’s
functions should be inhibited and those instruments subject to interference should be moved
away. You can view the EMC information