Software User Manual - Argos 3D-P320
Last change: 2 February 2016
Version 0.14
© Bluetechnix 2016
Page 62 | 70
Firmware
Version
Status Release
date
Changes
o
Chapter 4.15.3
Change
DeviceInfo
register content
Add register
Revision
to store overall camera revision
Switch off RGB LED by default (near color sensor)
o
Register
RgbLedColor
Rework Status Indicator LED
o
Chapter 3.7
0.12.0
X-
Grade
Feb 2015
New firmware for TIM
Rework wiggling correction (wiggling correction file is needed)
Add 2x2 average filter
Table 6-1: Overview Argos 3D-P320 firmware changes
Note
Please refer to our support site for additional information about product changes.
6.2
Anomalies
Applies to
Date
Description
0.2.2
2014 11 13 Factory default reset resets DistOffsetX registers to zero. Offset
calibration must be executed for each modulation frequency.
0.3.1
2014 11 13 If your Argos 3D-P320 was delivered with firmware version 0.2.2
originally, factory default reset resets DistOffsetX registers to zero.
Please contact Bluetechnix support for instructions how to configure
factory default settings.
<= 0.3.1
2015 04 20 Illumination not turned on after boot in rare cases
0.4.2
2015 07 08 FPPN calibration data is applied partly incorrect on distance data
0.5.0
2015 08 27 Color sensor stream problem may occur on changing parameters
<= 0.5.0
2015 08 27 TIM temperature readout problem (Values in register and header
freeze)
0.9.0
2015 10 13 If the time between two snapshots is too short, the camera switches
to manual trigger mode without finishing the snapshot
Table 6-2: Firmware anomalies