Blue-Ray Biotech MiniTurbo Скачать руководство пользователя страница 16

MiniTurbo Operation Manual 

                                                                                                
                                                                                              

12 

6.  Troubleshooting 

6.1   General Problems 

 

 

6.2   Error Messages 

The instrument will show an error message on the main screen and stop working until the problem 

has been solved. Please see the table below. 

Message 

Cause 

Action 

Er01- Heater overheat   

Heater over 120

°

Reboot the unit. 

Er02- Heater cannot reach 

the setting temperature. 

Faulty heater. 

Reboot the unit. 

Er03- Heater has lost 

temperature accuracy 

The heater temperature has 

been over ±3

°

C for 30 seconds. 

Reboot the unit. 

Er04- Heater temperature 

sensor error. 

Heater temperature sensor 

problem. 

Reboot the unit. 

Er05- Block temperature 

sensor error. 

Block temperature sensor 

problem. 

Reboot the unit. 

Problem 

Cause 

Action 

The display remains 

off even when power is 

switched on. 

Power is not reaching the system. 

Check power source voltage. 

The power cord is not properly 

plugged into the socket. 

Reconnect the power cord. 

Faulty power switch. 

Return the unit for service. 

Lid will not open or 

close. 

Foreign object between lid heater 

and sample block. 

Remove the foreign object or 

matter. 

Faulty lid lock mechanism. 

Return the unit for service. 

The display goes off. 

Faulty backlight. 

Return the unit for service. 

Faulty LCD panel 

Return the unit for service. 

Cycle time is too long. 

Operating environment 

temperature may be unsuitable. 

Make sure the temperature of the 

operating environment is between 

15 and 30°C. 

The electronic cooling element 

may be damaged or aged. 

Return the unit for service. 

Faulty temperature sensor. 

Return the unit for service. 

Lid heater does not 

work. 

Lid heater is set to off. 

Check Lid Temperature setting in 

protocol. 

Lid sensor problem. 

Return the unit for service. 

Содержание MiniTurbo

Страница 1: ...MiniTurbo Thermal Cycler TCMI 0810 Operation Manual Ver 1 0 Copyright 2021 Blue Ray Biotech Corp All Rights Reserved...

Страница 2: ......

Страница 3: ...eration 7 4 2 Lid Opening Closing 7 4 3 Protocol Chip 8 4 4 Heated Lid 9 4 5 Loading the Reaction Vessel 9 4 6 Main Screen 10 4 7 Start Running 10 5 Preventative Maintenance 11 5 1 Cleaning the Unit 1...

Страница 4: ......

Страница 5: ...ce in direct sunlight 3 Choose a flat stable surface capable of supporting the weight of the device 4 Make sure the power source conforms to the required power supply specifications 5 To avoid electri...

Страница 6: ...rovides the same precise test results as a normal sized instrument It is a well designed reliable and convenient thermal cycler that can be used anywhere 2 1 Features The instrument is compact in size...

Страница 7: ...un Start Button Starts the protocol running or switches a routine Lid The heated lid is designed to prevent condensation inside the reaction vessels during PCR temperature cycling Lid Open Button Pres...

Страница 8: ...heat conduction It will also help prevent the leakage of sample vapor caused by weak vessel caps or poor sealing Sample Block The sample block holds the reaction vessels Protocol Chip Port The program...

Страница 9: ...al Description 5 Figure 3 Rear view Table 3 Detailed description for rear view Name Function Power Cable Socket Power cable socket compartment Power Switch Power On Off switch Power Cable Socket Power...

Страница 10: ...ms are included MiniTurbo unit x 1 Programmable Chip x 1 MiniTurbo Writer x 1 MiniTurbo Writer USB connector x 1 USB Flash Drive x 1 AC power adapter x 1 AC power cord x 1 If there are any items missi...

Страница 11: ...of the device Turn on the device by using the power switch at the rear of the unit The main screen will light up for 2 seconds and there will be a beep This indicates that the power is on Switch off...

Страница 12: ...Make sure the lid is right down as show in Figure 5 Figure 5 Closing the lid 4 3 Protocol Chip The protocol chip has stored protocols and there is no need to enter any protocols by hand This prevents...

Страница 13: ...ture unchanged during the heating and cooling cycles The heater platen also applies pressure to the caps or sealing film on the vessels to prevent vapor loss and cross contamination between samples 4...

Страница 14: ...n the protocol While the protocol is running the start button acts as a change over switch that toggles the main screen display between remaining time and heater temperature If necessary the start but...

Страница 15: ...h is to remove dust and other residue that comes with normal operation of the unit A vacuum cleaner can be used to remove dust from the air vents 5 2 Cleaning the Heated Lid Make sure the MiniTurbo is...

Страница 16: ...hen power is switched on Power is not reaching the system Check power source voltage The power cord is not properly plugged into the socket Reconnect the power cord Faulty power switch Return the unit...

Страница 17: ...even when Protocol Chip is correctly inserted Faulty Protocol Chip Return the Protocol Chip for service Faulty protocol running port Return the unit for service Protocol Chip is not properly plugged i...

Страница 18: ...ating Rate 5 5 C sec Max Cooling Rate 3 4 C sec Temperature Accuracy 0 4 C Temperature Uniformity Across Block 0 4 C General Display A group of 6 digits display Heated Lid Yes fixed 105 C pre heat to...

Страница 19: ...C Declaration of Conformity Product Name Thermal Cycler Model Names MiniTurbo All models comply with the following European standards EMC EN 61326 Group 1 Class A Safety EN 61010 1 and EN 61010 2 101...