4.
Retrieving the contents:
a.
Repeat step 1.
b.
In Step 2, carefully remove the samples/contents from the cold chamber, while ensuring that
the sensor in the cold chamber is not damaged by the contents during removal.
c.
Repeat step 3 to close the device lid.
NOTE!
Ensure the contents are placed in a manner such that they are not at risk of moving around during transport and
leading to breakage. Use of zip-lock bags to contain any glass vials is recommended. Vial-tray (accessory) needs to
be pre-loaded before depositing in the cold-chamber.
Before the contents are transferred from a stationary refrigerator onto Emvólio, it is recommended that the system
be pre-cooled to the desired temperature. This may take about 10- 15 minutes depending on the ambient
temperature of the surroundings. This p
rocess will ensure there’s no freeze
-thaw cycle for the contents.
3.2.3
Setting chamber temperature
The following steps have to be followed to set the desired chamber temperature:
•
Ensure device is in ON condition.
•
Press and hold SET BUTTON for 2 seconds to enter temperature selection mode.
•
Use (+) and (-) temperature adjustment buttons to adjust chamber temperature, in steps of 1°C.
•
Press OK button to confirm and set cold chamber temperature.
Set Button
Temperature
adjustment buttons