bk3500 User Guide
(16-01410-EN-04)
Getting Started
25
Critical Low Battery
The setting for critical low battery is 3 minutes. When the battery reaches the critical
level, the system shuts down immediately.
When the system is plugged in, the display on the battery compartment is lit.
NOTE:
The time settings for low and critical low battery can be changed. Contact
your BK service representative.
Battery Life
Typical vs.
continuous
use
With typical use (which includes imaging and freezing), a fully charged battery
provides approximately 2 hours of operation. With a fully charged battery pack, you
can scan continuously for approximately 1 hour 20 minutes.
Battery
lifespan
When a battery reaches the end of its lifespan, the replace battery message appears
on the monitor at start-up.
On average, a battery will need to be replaced after approximately 300 recharge and
discharge cycles – the battery lifespan. Contact your BK service representative for
replacement of batteries.
For information about the disposal of depleted batteries, see page 40.