98
5.2
Voltage Calibration
Follow the step by step instructions below for voltage calibration.
1.
Power off the supply and connect its output terminal to a Digital
Volt Meter (DVM), as shown in the figure below. Then set the
meter for DC voltage measurement.
DVM
DC Voltage measurment
Voltage Calibration
DS30052
+
-
2.
Turn on the power supply’s output.
3.
Press
in the third page of Menu Setting to enter
CALIBRATION page. Users can enter the password and then access to
the calibration procedures.
Содержание XLN15010
Страница 1: ...Model XLN15010 XLN30052 XLN60026 High Power Programmable DC Power Supply USER MANUAL...
Страница 20: ...17 voltage is within the specified range...
Страница 111: ...108...
Страница 113: ...110...
Страница 114: ...111 22820 Savi Ranch Parkway Yorba Linda CA 92887 www bkprecision com 2013 B K Precision Corp Printed in Taiwan v042413...