Operations
38
CH01
Rset = 7500.0
Ω
Vmax = 82 V
Vmin = 0 V
RLC R= 7500.0
Ω
RLC L = 0 uH
RLC C = 10 uF
Rset: set the impedance value (
𝑅
𝐿
)
Vmax: set maximum voltage pass/fail limit for Automatic test mode
Vmin: set minimum voltage pass/fail limit for Automatic test mode
RLC R: set the series resistance value (
𝑅
𝑠
)
RLC L: set the series inductance value (
𝐿
𝑠
)
RLC C: set the parallel capacitance value (
𝐶
𝐿
)
4.1.5.1 Setting CV,CC,CR, CW,CZ Mode
To set up an operation mode from the front panel:
1. Power on the electronic load. Self-test
2. Press the
keys to select the channel to be edited, such as channel 1. CH01 CC OFF
3. Press the the
key to enter the channel setup menu.
4. Press the the
key to enter the
Mode
menu.
5. Press the the
key to select CV, CC, CR, CW, or CZ mode.
– Press the the
key to confirm.
6. Press the the
key to select the
Range
menu.
– Press the the
key to confirm
7. Press the the
key to select
Low Range
or
High Range
.
– Press the the
key to confirm
8. Press the the
key to select the voltage setting
Vset
, current setting
Iset
, resistance setting
Rset
, power setting,
or impedance setting.
– Press the the
key to confirm.
9. Press the
key to select the rise slope setting
∫
. Input the value and Press the
– Input the value and Press the the
key to confirm. (CC mode only)
10. Press the the
key to select the fall slope setting
∫
.
– Input the value and Press the the
key to confirm. (CC mode only)
11. Press the the
key to exit.
12. Press the the
key to turn on the load’s input.