Load Connection
31
3.6 PC Control Connection
The MDL4U Series DC Electronic Load can achieve remote control via USB, RS-232, LAN, or GPIB interface, but only
one interface can be used at a time.
To choose the interface:
1. Connect communication cable before powering on.
– Do not hot plug, as it may damage the communication interface of the electronic load.
2. Power on the electronic load.
3. Select the channel number
via keys.
4. Press the
+
to enter the
System
menu.
5. Select
Communication
and select the desired interface.
– Press Enter to confirm.
6. The display will return to the main menu.
Note:
When using RS232, GPIB, or LAN the commuincation settings must be set. To do so select RS232 Set,
GPIB Address, or Ethernet Set from the system menu.
See the
for full listing of menus and submenus applicable to all other remote interfaces.