27
Example
CURR:SLEW 3
Related Commands
[SOURce:]CURRent:SLEW?
Query Response
<NR2>
7.11 [SOURce:]CURRent:SLEW:RISE <NRf+>
This command sets the slew rate of the current for positive going transitions.
Command Syntax
[SOURce:]CURRent:SLEW:RISE <NRf+>
Parameters
MIN MAX - MAXimum - MINimum
Unit
A/uS
*RST Value
MAXimum
Example
CURR:SLEW RISE 3
Query Syntax
[SOURce:]CURRent:SLEW:RISE?
Query Response
<NR2
7.12 [SOURce:]CURRent:SLEW:FALL <NRf+>
This command sets the slew rate of the current for negative going transitions.
Command Syntax
[SOURce:]CURRent:SLEW:FALL <NRf+>
Parameters
MIN MAX - MAXimum - MINimum
Unit
A/uS
*RST Value
MAXimum
Example
CURR:SLEW FALL 3
Query Syntax
[SOURce:]CURRent:SLEW:FALL?
Query Response
<NR2>
7.13 [SOURce:]CURRent:PROTection[:LEVel] <NRf+>
This command sets the current protection level. If the input current exceeds the current protection level for the time
specified by CURR:PROT:DEL, the input is turned off.
NOTE:
Use CURR:PROT:DEL to prevent momentary current limit conditions caused by programmed changes from
tripping the overcurrent protection.
Command Syntax
[SOURce:]CURRent:PROTection[:LEVel] <NRf+>
Parameters
MIN MAX - MINimum - MAXimum
Unit
A