Utility Menu
218
Note:
23.8
Start Self Cal
The self-calibration program can quickly calibrate the oscilloscope to reach the best working state and the most precise
measurement. It is recommended to perform a self-calibration if the change of ambient temperature is more than 5
℃.
Make sure the oscilloscope has been warmed up or operated for more than 30 minutes before the self-calibration.
To perform a self-calibration :
1.
Disconnect all inputs.
2.
Press the
Utility
key on the front panel to enter the Utility menu.
3.
Touch
Start Self Cal
and it will pop up the following dialog. Select Continue to start the self-cal program.
4.
The oscilloscope will not respond to any operation until the self-cal is finished. After the self-cal is completed,
touch the screen or press any button to exit.
23.9
Power On Line
To set the power on line:
1.
Press the
Utility
key on the front panel.
2.
Press the
Power On Line
softkey to select
Enable
or
Disable
.
•
Enable
: when the oscilloscope is energized, the oscilloscope will start automatically by itself.
•
Disable
: when the oscilloscope is energized; then; press the power key at the lower-left corner of the front panel
to start the oscilloscope.