6.3 Calibration
Calibration is a procedure that ensures that the power supply will work properly,
with parameters specified within the Specifications section.
Before initiating the calibration procedure, the following conditions must be assured:
•
Disconnect any loads connected to the power supply and turn it on
•
Let the power supply be turned on for 1 hour with no loads connected before
you start the calibration procedure
•
Calibration ambient temperature must be 25
°
C
•
Ambient relative humidity must be less than 80%
To calibrate the power supply, complete the following steps:
1.
Turn on the power supply in calibrating mode
2.
Zero voltage calibration
3.
Full voltage calibration
4.
Zero current calibration
5.
Full current calibration
6.
Leaving the Calibration menu
Figure 11 - Location of Adjustments