19
3. SAVE<CR>
The programmed values of the output current and voltage are saved. If the
power supply is restarted, the values of the output current and voltage will
be those previously saved.
4. VOLT?<CR>
This command returns the voltage measured to the output terminals of the
power supply: xx.xxV<CR>
The voltage value returned is the same as the one displayed on the voltage
display.
5. CURR?<CR>
This command returns the current measured to the output terminals of the
power supply: x.xxxA<CR>
The current value returned is the same as the one displayed on the current
display.
6. STAT?<CR>
The power supply responds to this command with one of the following
messages:
CV<CR> if the power supply is in Constant Voltage mode
CC<CR> if the power supply is in Constant Current mode
5.3
Application Software
Remotely control your 1737 using B&K Precision’s application software, available for
downloada simple data logging function to store and log data to a text or CSV file. RS232
commands can also be sent through the software.