:ANAlyzer#:THReshold
Syntax:
:ANAlyzer#:THReshold <threshold>
Defines the threshold, in volts, of the input comparator.
This parameter has no effect if the input is configured to differential mode.
Example command:
ANA0:THR -0.2
:ANAlyzer#:THReshold?
Syntax:
:ANAlyzer#:THReshold?
Returns the currently programmed threshold level, in volts.
Example query:
ANA0:THR?
Example response:
-200e-3
:ANAlyzer#:MODE
Syntax:
:ANAlyzer#:MODE <SINGle|DIFFerential>
Selects between single-ended input mode (
SINGle
) or differential input mode (
DIFFerential
).
In single-ended input mode, the programmed threshold voltage is used.
In differential input mode, the threshold voltage is not used; instead, the complement analyzer input
is used as a reference.
Example command:
ANA0:MODE SING
:ANAlyzer#:MODE?
Syntax:
:ANAlyzer#:MODE?
Returns the currently configured input mode.
Example query:
ANA0:MODE?
Example response:
SINGle
:ANAlyzer#:SLOT?
Syntax:
:ANAlyzer#:SLOT?
Returns the slot number where the module is mounted which holds this analyzer input. The first
front-end module slot has number 1.
Example query:
ANA1:SLOT?
Example response:
2
BIT-3000 User Manual
49