Setting the Full Scale Power
For element-based sensors, enter the power rating
listed on the forward element before making any mea-
surements. Also, check that the reflected element’s
rating is 10% of the forward element’s.
The Bird Site Analyzer will automatically set the full
scale power for sensors that do not use elements, such
as the Bird 5011 or 5012.
Zeroing Bird Power Sensors
Over time, the sensor’s “zero value” (reading with no
applied RF power) can drift, making all readings inac-
curate by this value. For example, if the zero value is –
2 µW, measuring a 5 mW signal will give a reading of
Press from
Measure
Power Mode
Scroll to Full
Scale
Enter the
forward
element
power
Offset
FullScale
0.00 dB
1000.00
Watts
Fwd
0.000 KWatts
1.20
Recall
Setups
Measure Power Mode
Configuration Screen
CAUTION
When using a Bird 5011 or 5011-EF, do not exceed 2
W average or 125 W peak power for 5 µs. Doing so
will render the sensor inoperative.