PC/Rack Mount SignalHawk with ASL
148
Update PC SignalHawk Firmware
NOTE
There are two different firmware files: one containing the DSP image and the other containing
the FPGA image.
1.
Navigate to Bird Technologies Group>PC SignalHawk in the Start menu.
2.
Click on Firmware Update Wizard.
NOTE
The wizard will launch and provide step-by-step instructions
3.
Select the DSP and FPGA images to be updated.
DSP and FPGA Image Selection
4.
Connect the unit to the PC via USB cable.
5.
Turn on the SignalHawk.
6.
Select “Install the software automatically...” if the “Found New Hardware Wizard” is
presented.
NOTE
Both program wizards will run concurrently. See
7.
Click “Next.”