Settings
28
Freq & Span Menu
Figure
19
Freq & Span Menu
In the Freq & Span Menu the range of frequencies to sweep are specified. The
SignalHawk can sweep frequencies between 100 kHz and 3.6 GHz.
Frequencies can be selected at spans from 1 kHz up to the entire range of the
instrument. These frequencies can be set to sweep by one of the following
methods:
Set the Center Freq and Freq Span.
Set the Start Freq and Stop Freq.
Select the Full Span of the instrument.
Select a band from the Freq List.
The most common of these methods is to set the center frequency and a
frequency span (a range) and let SignalHawk calculate the start and stop
frequencies for the sweep. Reducing the span, by default (with RBW and VBW,
both on auto mode) will usually speed up the sweep, provide more detail, and
lower the noise floor. Unless specified otherwise, set the center frequency to
the center of the signal being measured.
Содержание SignalHawk SH-36S-PC
Страница 26: ...SignalHawk PC and Rack Mount Models 11 Figure 7 Select DSP Device List ...
Страница 67: ...Measurements 52 Figure 41 Example Channel Power Integration Bandwidth STEP 2 STEP 3 ...
Страница 75: ...Measurements 60 Figure 49 Example Carrier to Interference Ratio Carrier Off STEP 4 Measurement Settings ...
Страница 116: ...SignalHawk PC and Rack Mount Models 101 Figure 75 Menu Map BW Sweep Menu All Measurements ...
Страница 117: ...Menu Maps 102 Figure 76 Menu Map Amplitude All Measurements ...
Страница 118: ...SignalHawk PC and Rack Mount Models 103 Figure 77 Menu Map Measurement All Measurements ...
Страница 119: ...Menu Maps 104 Figure 78 Menu Map Mark Limit Menu All Measurements ...
Страница 120: ...SignalHawk PC and Rack Mount Models 105 Figure 79 Menu Map File Help All Measurements ...
Страница 121: ...Menu Maps 106 Setup Function Menu Maps Figure 80 Map Setup Mode Main Screen ...