12
Chapter 4
Specifications
5012D Specifications
Sensor Characteristics
Frequency Range
350 MHz to 4 GHz
RF Power Range
0.15 W to 150 W average, 4W to 400
W peak
Maximum Power
See
Figure
8 on page
14
.
Impedance, Nominal
50 ohms
Insertion Loss, Max:
0.35 – 1 GHz
1 – 4 GHz
0.05 dB
0.1 dB
Input VSWR, Max:
0.35 – 2.5 GHz
2.5 – 4 GHz
1.05
1.10
Directivity, Min:
0.35 – 3 GHz
3 – 4 GHz
30 dB
28 dB
RF Connectors
N Female
Interface:
DPM
PC Serial Port
PC USB Port
Male DB-9, EIA-232, 9600 Baud, no
parity, 8 data bits, 1 stop bit
Female DB-9, EIA-232, 9600 Baud,
no parity, 8 data bits, 1 stop bit
USB 1.1 interface
Power Supply:
DPM
USB Port
DC Connector
From host instrument via cable
Less than one low-power USB load
7 – 18 Vdc, < 100 mA
Data Logging
In VPM Software
Содержание 5012D
Страница 24: ...Wideband Power Sensors 16 Figure 9 5012D WPS...
Страница 29: ...Specifications 21 Figure 12 5016D WPS...
Страница 34: ...Wideband Power Sensors 26 Figure 15 5017D WPS...
Страница 39: ...Specifications 31 Figure 18 5018D WPS...
Страница 44: ...Wideband Power Sensors 36 Figure 21 5019D WPS...
Страница 45: ...Specifications 37...