6.3 Pulse and Control Parameters
Basic rate
30...(1)...
60
...(1)...88...(2)...122...(3)...140...(5)...180 ppm
Night rate
Off
, 30...(1)...60...(1)...88...(2)...122...(3)...140...(5)...180 ppm
Rate Hysteresis
Off
; -5...(5)...-50 bpm
Repetitive Hysteresis
Off;
1…(1)…10
Scan Hysteresis
Off
: 1…(1)…10
Upper Tracking Rate (UTR)
100; 110; 120;
130
; 140; 160; 185 ppm
UTR Response
2:1; WRL (automatic selection)
Rate Limitation
2,3,4
190…220 ppm
Dynamic AV Delay (Dual chamber only)
low
; medium; high; individual; fixed
AV Delay Values (Dual chamber only)
15; 50; 75; 100; 120...(10)...200; 225; 250; 300 ms
(Programmable in 5 ranges)
2
The corresponding intervals t correlate with the rates f by the formula t = 60.000
/ f (t in ms, f in ppm).
3
In the event of electronic defect.
4
Rate Limitation changes as the Pacemaker approaches End of Service. The
Rate Limitation is nominally 190 ppm at Beginning of Service (BOS) and can
reach 220 ppm at End of Service (EOS) due to battery depletion.