Страница 2: ...elines 9 Pre rinsing 9 Pipetting Liquids of Varying Densities 9 Temperature Considerations 9 Reverse Pipetting 10 Troubleshooting Troubleshooting and Repairs 11 Sample Splash 12 Leaks Inaccuracy Abnormal Stroke 12 Service Calibration and Repair 12 Maintenance Autoclaving 13 Tip Ejector Arm Removal 13 Storage 13 Acids and Corrosives 13 Parts 13 Accessories Pipette Carousel 14 Replacement Parts 14 W...
Страница 3: ...ed and are undamaged xPIPETTE xPIPETTE User Guide Calibration Certificate GLP Compliance The unique pipette serial number is engraved on the body of the pipette underneath the handle It provides a unique identification to to help fulfill GLP requirements Calibration certificates are also included for further traceability Description Connecting Nut Shaft Stainless Steel Tip Ejector Arm Digital Volu...
Страница 4: ... 04 0 20 0 10 10 0 15 0 05 0 20 0 10 20 0 2 0 06 0 20 0 10 100 μL with xTIP4 200 μL tips P N 63300163 10 0 35 0 10 0 80 0 30 50 0 4 0 12 0 80 0 30 100 0 8 0 15 0 80 0 30 200 μL with xTIP4 200 μL tips P N 63300164 20 0 5 0 20 1 60 0 60 100 0 8 0 25 1 60 0 60 200 1 6 0 30 1 60 0 60 300μL with xTIP4 300 μL tips P N 63300327 30 0 75 0 30 2 40 0 90 150 1 2 0 375 2 40 0 90 300 2 4 0 45 2 40 0 90 1000μL ...
Страница 5: ...with xTIP4 20 μL tips P N 63305171 63305175 1 0 15 0 08 0 30 0 16 5 0 15 0 15 0 30 0 30 10 0 20 0 10 0 40 0 20 200 μL with xTIP4 200 μL tips P N 63305172 63305176 2 0 50 0 24 1 00 0 48 10 0 80 0 25 1 60 0 50 20 1 60 0 60 3 20 1 20 300 μL with xTIP4 300 μL tips P N 63305173 63305177 10 0 75 0 10 1 50 0 20 50 1 20 0 12 2 40 0 30 100 2 40 0 15 4 80 0 30 Specifications are subject to change without no...
Страница 6: ...All digits black whole μL 1000 μL Red mL Black tenths hundreds of mL 3 To eliminate errors due to mechanical backlash when setting the desired volume first turn the knob 1 3 turn above the desired volume Then turn the knob slowly clockwise until the desired volume is displayed 4 Turn the volume lock clockwise see diagram above to prevent any accidental changes to the volume setting Example volumes...
Страница 7: ... Adjustable Range μL Recommended Volume Range μL Increment μL 2 0 2 0 1 2 0 002 10 0 10 0 5 10 0 02 20 0 20 2 20 0 02 100 0 100 10 100 0 2 200 0 200 20 200 0 2 300 0 300 30 300 0 5 1000 0 1000 100 1000 2 0 Pipette Size μL Adjustable Range μL Recommended Volume Range μL Increment μL 10 0 10 0 5 10 0 02 20 0 20 2 20 0 02 200 0 200 20 200 0 02 300 0 300 30 300 0 5 xPIPETTE Manual ...
Страница 8: ...d on page 7 2 Attach a new tip Press the shaft into the tip to make a good seal 3 Press the plunger button to the FIRST STOP and hold it in this position The magnetic assist will help you sense and hold this position 4 Holding the pipette vertically place the tip into the sam ple to the proper depth and relax your thumb pressure on the plunger The light piston spring will move the piston upward as...
Страница 9: ...p Pre Rinsing Recommended Some solutions may leave a film on the inside tip wall This film remains relatively constant in successive pipetting with the same tip so excellent precision can be obtained by refilling the tip and using the refilled volume as the sample Successive samples from this same tip will exhibit good reproducibility Temperature Considerations Warm or cold liquids can be measured...
Страница 10: ...or more viscous liquids is reverse pipetting The operating sequence is reversed 1 Mount a disposable tip on the pipette shaft nozzle 2 Press the plunger button fully to the SECOND STOP 3 Immerse the tip in liquid and allow the button to return slowly to the resting position Wait a moment for the liquid column in the tip to reach equilibrium 4 Wipe any excess liquid from the outside of the tip with...
Страница 11: ...Do not use instrument Re turn instrument to supplier Pipette is inaccurate Improper repair or assembly See page 12 Loose connecting nut Tighten the connecting nut Pipette is not precise Loose connecting nut Tighten the connecting nut Incorrect operator technique Operator training Damaged piston Return pipette to supplier Damaged shaft Replace the shaft Worn O ring or seal Replace both parts Tips F...
Страница 12: ...vice 3 Worn seal and or o ring All models up to 1000 μL incorporate a polyethelyne seal and o ring Examine the seal and o ring replacing them as necessary Pull off the old seal and o ring position the new seal and o ring on the piston assembly as shown in the drawings on page 13 and reassemble the pipette 4 Forming the seal xPIPETTE uses a small amount of specialized high quality grease applied to...
Страница 13: ...ents thoroughly and reassemble Extensive contact with corrosive fumes may result in premature seal wear and damage to the piston Exposure of internal components to corrosive fumes can be reduced by using tips with aerosol barrier filters Parts 20 μL 200 μL 1000 μL A A A B B B C C C Liquid End Ejector Arm A Piston As B Seal Reta C Seal 20 μL 200 μL 1000 μL A A A B B B C C C Liquid End Ejector Arm A...
Страница 14: ...7 pipettes Part Number Description 30539600 2μL Shaft 30539517 10μL Shaft 30539599 20μL Shaft 30539516 100μL Shaft 30539598 200μL Shaft 30539601 300μL Shaft 30539515 1000μL Shaft 17001797 Ejector Arm 20μL 17001799 Ejector Arm 200μL 17007556 Ejector Arm 300μL 17001793 Ejector Arm 1000μL 17014318 O Ring 010 ID 2μL C pg 13 17014689 O Ring 010 ID 10μL C pg 13 17014320 O Ring 010 ID 20μL C pg 13 170142...
Страница 15: ...ix uTIP xPIPETTE xTIP4 FlexFit StarStop X Resin Blade Tip Eject and CleanPak are trademarks of Biotix inc All other brands and names contained are the property of their respective owners as found at biotix com trademarks BIOTIX BIOTIX com com ...