Chapter 4: Preventive Maintenance
Removing the Internal Tubes and Injector Heads
Take a moment to identify the components described in this section:
Figure 20: Internal components for the injection system
Important!
When reinstalling the
internal dispense tubes, be sure to
align the tubing ports with the injector
heads as shown in this diagram. Look
for the
SYRINGE 1
and
SYRINGE 2
labels on the instrument’s rear panel.
SYRINGE 1 SYRINGE 2
Internal tubes
(2, black) for
delivering fluid
to the injector
heads
Injector heads
(2, white) for dispensing
fluid into microwells
Cable clamp
for holding the
tubes in place
Tubing ports (2) for connecting
the two internal dispense tubes to
the two external dispense tubes