9. Discard the used gloves and cloths using a biohazard trash bag and an approved
biohazard container.
Clean the Tubing and Injectors
Perform the procedures under
Clean the Dispense Tubes and Injectors
on page
Decontaminate the Tip Priming Trough and Priming Plate
1. Remove the tip priming trough from the instrument’s microplate carrier.
2. Wash the tip priming trough and priming plate in hot, soapy water. Use a small
brush or cloth to clean the corners of the trough and plate.
3. To decontaminate, soak the trough and plate in a container of 0.5% sodium
hypochlorite or 70% isopropyl alcohol for at least 20 minutes.
o
If decontaminating in a bleach solution, thoroughly rinse the trough and plate
with DI water.
o
If decontaminating with alcohol, let the trough and plate air dry.
4. Discard the used gloves and cloths using a biohazard trash bag and an approved
biohazard container.
Alternate Procedure
If you are unable to prime the system due to an equipment failure, decontaminate the
instrument and the dispense module as follows:
1. Perform the procedures under
Clean the Dispense Tubes and Injectors
on page
.
2. Prepare an aqueous solution of 0.50% sodium hypochlorite (bleach). If the effects
of bleach are a concern, 70% isopropyl alcohol may be used.
Check the percent NaClO of the bleach you are using. Commercial bleach is
typically 10.0% NaClO; prepare a 1:20 dilution. Household bleach is
typically 5.0% NaClO; prepare a 1:10 dilution.
3. Slide the microplate carrier out of the instrument.
4. Wet a cloth or paper towel with the bleach solution or alcohol, and then
thoroughly wring it out so that liquid does not drip from it
.
5. Use the cloth to wipe:
o
All exterior surfaces of the instrument
o
All surfaces of the plate carrier
o
The exposed surfaces of the dispense module, including the syringe valves
6. Remove the tubing and the syringes from the dispense module and soak them in
the bleach or alcohol solution. Wait for 20 minutes.
70 | Chapter 6: As-Needed Maintenance
BioTek Instruments, Inc.
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...