Clean Exposed Surfaces
Exposed surfaces may be cleaned (not decontaminated) with a cloth moistened (not
soaked) with water or water and a mild detergent.
You will need:
l
Deionized or distilled water
l
Clean, lint-free cotton cloths or paper towels
l
Mild detergent (optional)
Procedure:
1.
Important!
Turn off and unplug the instrument.
2. Wet a cloth or paper towel with water, or with water and mild detergent, and then
thoroughly wring it out so that liquid does not drip from it
.
3. Wipe the plate carrier and all exposed surfaces of the instrument.
4. Wipe all exposed surfaces of the dispense module (if used).
5. Wipe all exposed surfaces of the gas controller module (if used).
6. If detergent was used, wipe all surfaces with a cloth moistened (not soaked) with
water.
7. Use a clean, dry cloth to dry all wet surfaces.
Models with injectors:
If the tip priming trough overflows or other spills occur inside
the instrument, wipe the carrier and the surface beneath the carrier with a dry cotton
cloth. The internal chamber and probes are not customer-accessible. If overflow is
significant, contact BioTek’s Technical Assistance Center.
Synergy H1 Operator's Manual
Clean Exposed Surfaces | 57
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...