Gen5 Software
Gen5 supports all Synergy H1 models. Use Gen5 to control the reader, the dispense module
(if equipped), and the BioStack (if equipped); perform data reduction and analysis on the
measurement values; print or export results; and more. This section provides brief
instructions for working with Gen5 to create protocols and experiments and read plates.
Refer to the
Gen5 Getting Started Guide
and the Help system for more information.
Define the Filter Cube in Gen5 and on the Reader
As described on page
, most Synergy H1 models are delivered with a filter cube
installed, and the reader’s onboard software is configured with the filter and mirror values
and their locations in that filter cube. When Gen5 communicates with the reader, it “asks”
for this information and then stores the values in a Filter Cube table.
Important!
It is critical that the values in Gen5 and the reader's
software exactly match the contents of the installed filter cube.
If you exchange or modify the filter cube, you must update the Gen5 Filter Cube table and
send the information to the reader:
1. From the Gen5 main view, select
System > Instrument Configuration
. Highlight
the
Synergy H1
, click
View/Modify
and then click
Setup
.
2. If this is a new filter cube, enter a unique name to identify the cube and then enter a
name for Filter Set 1.
3.
If applicable
, check the Fluorescence Polarization Cube box.
4. Define/modify settings for the excitation and emission filters:
l
Select
Band Pass
,
Long Pass
, or
Short Pass
, as appropriate for each filter
type.
Band Pass
a standard interference filter with a defined central
wavelength and bandwidth
Long Pass
cutoff filter that transmits longer wavelengths and block
shorter wavelengths
Short Pass
cutoff filter that transmits shorter wavelengths and
block longer wavelengths
l
Select
PLUG
to indicate the presence of a plug.
l
Select
HOLE
to indicate an empty location.
Synergy H1 Operator's Manual
Gen5 Software | 35
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...