Figure 2-4: Dispense module components
12. Remove the two syringes from their protective boxes. They are identical and
interchangeable. Each syringe should already be assembled in one piece, but if for
some reason there are two separate pieces, assemble them now: insert the white tip
of the syringe plunger into the barrel of the syringe and gently push it all the way into
the barrel.
13. Install the syringes, referring to
on page
:
l
Hold the syringe vertically with the threaded end at the top.
l
Screw the top of the syringe into the bottom of the syringe valve. Finger-tighten
only.
l
Carefully pull down the bottom of the syringe until it rests inside the hole in the
bracket.
l
Pass a thumbscrew up through this hole and thread it into the bottom of the
syringe. Hold the syringe to prevent it from rotating while tightening the
thumbscrew. Finger-tighten only.
Synergy H1 Operator's Manual
7: Install the Dispense Module | 15
Содержание Synergy H1
Страница 1: ...Hybrid Multi Mode Microplate Reader Synergy H1 TM Operator s Manual ...
Страница 2: ......
Страница 26: ...BioTek Instruments Inc xxiv Preface ...
Страница 52: ...Figure 2 9 Packing the dispense module accessories 26 Chapter 2 Installation BioTek Instruments Inc ...
Страница 53: ...Figure 2 10 Packing the dispense module Synergy H1 Operator s Manual Repackaging and Shipping Instructions 27 ...
Страница 54: ...BioTek Instruments Inc 28 Chapter 2 Installation ...
Страница 68: ...BioTek Instruments Inc 42 Chapter 3 Getting Started ...
Страница 132: ...BioTek Instruments Inc 106 Chapter 7 Instrument Qualification Process ...
Страница 180: ...Figure C 1 Top view 152 Appendix C Instrument Dimensions for Robotic Interface BioTek Instruments Inc ...