Washer Qualification Liquid Tests | 251
OD of all wells on the plate. It is assumed that 1 mg = 1 µL of fluid for this
calculation.
Washer Qualification Test Materials
l
One new microplate per test to be performed:
Microplate Type
Flat-bottom 96-well plates, Corning® Costar #3590 or equivalent
Flat-bottom 384-well plates, Corning Costar #3574 or equivalent
Flat-bottom 48-well plates, Corning Costar #3548 or equivalent
Flat-bottom 24-well plates, Corning Costar #3524 or equivalent
Flat-bottom 12-well plates, Corning Costar #3513 or equivalent
Flat-bottom 6-well plates, Corning Costar #3516 or equivalent
l
Precision balance with minimum capacity of 100 g and readability of 0.001 g
resolution
l
Pipettes and graduated beakers
l
Microplate absorbance reader capable of dual wavelength reading at 630/450
nm
l
Liquid Test Worksheets
at the end of this chapter for recordingdata and
results
l
Deionized or distilled water
l
Test solutions
described below
n
These volumes are sufficient for performing the dispense tests and standard
and diagnostic Evacuation Efficiency tests with enough fluid remaining to re-
run a test, if necessary.
Test Solutions
Solution #1: Buffer Solution
Pipette 1 mL Tween 20
into 1 liter (1000
mL) of deionized or distilled water and
mix well.
or
Pipette 10 mL of BioTek Wetting Agent* into
1 liter of deionized or distilled water and
mix well.
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......