Syringe Dispense Step | 165
Enable
Pre-dispense
to normalize the tips before dispensing. Enter the volume and
number of cycles.
Select the
Columns
to dispense to (aka the Plate Map)
:
Each button represents a column.
Use the buttons and
Set
and
Clear
links: The buttons, which represent each
column, toggle on and off when selected. Toggle them on to dispense to the
column or off to skip the column.
l
Each button represents a
column
.
l
The number of active column buttons reflects the specified
Plate Type
, which should
match the plate you are dispensing to.
l
Learn about
for 384- and 1536-well plates.
Pump Delay
When dispensing highly viscous fluids, the tubing's check valves perform more slowly.
Delaying the syringe pump sufficiently to allow the specified amount of fluid to pass through
the check valves before being pumped into the syringe has been shown to improve dispense
accuracy.
Begin by setting the delay to 500 msec. Experiment with different settings to determine the
optimal value for your fluid.
Change the Syringe Dispenser Manifold
Changing the Syringe manifold requires two steps:
l
Physically changing the Syringe dispenser manifold:
Dispenser Component on page 62
l
Updating the instrument's manifold setting; as described below.
After physically changing the manifold, perform these steps to tell the instrument
which manifold is installed.
1. LHC: Select
Tools> Instrument Utilities> Syringe Dispenser
Touch screen: Select
Instrument> Syringe> Manifold
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......