Peri-pump Dispense Step | 155
Peri-pump Dispense Step
Add a Peri-pump dispense step to the protocol:
LHC
Touch screen
Click
P-Dispense
and define the parameters
Select
Add>P-Dispense
and
define the parameters
If applicable, select which Peri-pump to use, Primary or Secondary.
If applicable, enable the
dispenser.
Define the Dispense Step
Dispense Volume
: Enter the per well volume in microliters, up to 30,000/well. The
recommended values for each cassette type are:
o
1 µL cassette = 1 - 50 µL or 0.5 µL
o
5 µL cassette = 5 - 2500 µL
o
10 µL cassette = 10 - 3000 µL
Note:
When using the
RAD 8-to-1
bulk-dispensing cassette lower the dispense height to
ensure fluid hits the wells, i.e. select Advance Options and change the Z-axis.
Select the
Flow Rate
Cassette Type
1 μL
5 μL
10 μL
(μL/sec/tube)
Low
50
120
140
Medium
60
140
160
High
64
160
180
RAD
technology: Single-channel cassette flow rates match the standard cassettes and the
8-to-1 cassette matches the 10 µL flow rates.
When dispensing 0.5 µL with the 1 µL cassette the average flow rate values are:
1 μL Cassette
0.5 μL Dispense
(μL/sec/tube)
Low
52
Medium
54
High
56
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......