80 | Chapter 2: Installation
Verify the Peri-Pump Dispenser
1. Fill the Peri-pump’s supply bottle with approximately one liter of deionized
water.
LHC
Touch screen
1. Select
File > Open
.
2. Open the MultiFlo FX and
Maintenance Protocols folders, and
then open
P-xUL_CASS_RINSE.LHC
*
You
may
need to reset the Com Port.
3. When ready, click the
Run
button to
prime the tubing and manifold with
deionized water.
4. When finished, close the program.
1. Select
Quick
at the Home screen.
2. Select
Device Prime
and set the
Device
to the
Peri-pump
Primary
.
3. Press the
PRIME
button.
Repeat for Peri-pump Secondary, if
applicable.
4. When done, press the
Home
button.
*Verify both a standard cassette and a RAD technology cassette, if applicable.
If leaks were detected, tighten all tube fittings and re-run the program. If leaks are
still detected, contact BioTek’s Technical Assistance Center.
Verify the Syringe Dispensers
1. Fill the supply bottles with approximately one liter of deionized water, one for
each syringe pump.
LHC
Touch screen
1. Select
File > Open
.
2. Open the MultiFlo FX and
Maintenance Protocols folders, and
then open
S-DAY_RINSE_A&B.LHC
You
may
need to reset the Com
Port.
3. When ready, click the
Run
button
to prime the tubing and manifold
with deionized water.
4. When finished, close the program.
1. Select
Quick
at the Home screen.
2. Select
Device Prime
and set the
Device
to the
Syringe Both
.
3. Press the
Start
button.
4. When done, press the
Home
button.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......