The operating field must be filled with fluid preheated in the previous step, or with fluid at room
temperature. If required, the field can also be emptied to reach the appropriate level of fluid for
the treatment.
To change the working set temperature (43ºC by default) press the "-" and "+" buttons in the
"Temp. Set" box, which can be set between 20ºC and 48ºC. The temperature is controlled via
measurements from the irrigation probe; if this is not connected the corresponding alarm will
appear and the system will not work.
For information purposes, the temperature of the aspiration probe is also shown, although
experiments have shown that there is no fixed relationship between the entry and exit temperatures
of the liquid as this is a detail that varies greatly between one patient and another.
The emergency stop button can be used during any of the treatment processes and will activate
the stoppage and closing of the critical components (peristaltic pump, turbine and clamps).
The treatment will not restart automatically, the user must manually re-activate the
programme.
8.5
Treatment
In order to modify the set temperature, follow the same steps described above. In this step,
the default set temperature is 43ºC, although it can still reach 48ºC under the supervision of
the surgeon/anaesthetist. Once again, readings from the irrigation and aspiration probes are
shown.
If the
kit’s
probes are not connected to the monitoring cable and this in turn to the system, it
will not work.