Filling of the rack

Switch the motor on and homogenize 
the sample by gliding the „ball-bea-
ring head“ over the bags, using the 
handle.

Remove spills of extraction buffer immediately fi rst with a soft cloth wetted with 

water, then wipe dry - in particular when buffer entered the ball-bearing head. 

Occasionally grease the two gliding axes (5) and the nipple (6) at the ho-

mogenizing unit. Check  the holding-screws of the motor and the rack and 

tighten them if necessary. Replace the aluminium bars holding the bags 

when they look worn. They are made of „soft“ aluminium to prevent the ball-

bearing head from damage. 

Store the HOMEX at a dry place to avoid corrosion. When not in use for se-

veral weeks/months, additionally wipe all metal parts with a soft cloth wetted 

with oil (or by using an anti corrosion protection spray).

To change the rack, push the 

knobs (2) at the two fi xing-

screws and pull the screws 

out. Insert the rack and put the 

screws back in position.

To change or clean the ball-bearing head, open fi rst the screw on 

the side (3) and remove the cover-case. To remove the ball-bea-

ring head entirely, the head fi xation screw (4) has to be removed 

as well.

Insert 1 to 5 extraction bags in the 
open rack.

Close the rack by turning the two 
clamping-screws (1) to fi x the bags 
in the homogenization position.

Product Information

HOMEX 6 Instructions

Closing the rack

Homogenization

Maintenance and storage

Changing the rack