Alarm
No.
Alarm Message
Condition
Action
User Action
74
HMI A
COMMUNICATION
TIMEOUT
Communications between the
chamber and Door A HMI has been
interrupted.
If in cycle goes to
OFF mode.
Contact Service
Provider.
75
HMI B
COMMUNICATION
TIMEOUT
Communications between the
chamber and Door B HMI has been
interrupted.
If in cycle goes to
OFF mode.
Contact Service
Provider.
76
HVAC DAMPER
OPENING FAULT
HVAC OPTION 1 enabled. If time
expired and signal not received.
If in cycle goes to
OFF mode.
Investigate BMS system
and its signals. If OK
contact BMS supplier.
77
HVAC DAMPER
CLOSING FAULT
HVAC OPTION 1 enabled. If time
expired and signal not received.
If in cycle goes to
OFF mode.
Investigate BMS system
and its signals. If OK
contact BMS supplier.
78
BMS FAULT
HVAC OPTION 1 disabled. If time
expired and signal not received.
If in cycle goes to
OFF mode.
Investigate BMS system
and its signals. If OK
contact BMS supplier.
7.2
Fault Finding
Fault/Problem
Site Checks
Further Action
Door is difficult to open
Check for damage to door
and seal
Contact Bioquell Service
Provider
An alarm is displayed
Refer to Section 7.1 and
perform checks
suggested
Contact Bioquell Service
Provider
H
2
O
2
is detected outside
the chamber
Check the door seals are
not leaking
Contact Bioquell Service
Provider
HPV Generator will not
run a cycle
Check the connection of
the control cable between
the Generator and
Chamber
Contact Bioquell Service
Provider
Incorrect Password
Re-enter password in
accordance with User
Manual
Re-set password by
administrator
Contact person with
administration rights
Contact Bioquell Service
Provider