Operation Quick Start

1. Turn on the power.

2. Double click the icon

Q-Analyzer

for Qsep100

.

3. 

Q-Analyzer 

for Qsep100 software user interface:

4. Click “Connect” to link the instrument.

5. Cartridge and Buffer preparation: 

5-3. Alignment Marker preparation for cartridge:

Add ≥ 20 μl “Alignment Marker” into the 0.2 ml tube and then add 

10 

μl “Mineral Oil” on the top of  Alignment Marker.

5-

2. Add 1X Separation Buffer into “S” well 

and add diH

2

O into others.

* Buffer height should be equal to the  

groove of wells. 

5-1. Unpacking Cartridge:

Please follow the steps of “Unpacking guide” inside the cartridge 

box.

Packing list of Cartridge kit:

Cartridges
Alignment Marker
Mineral Oil
Dilution Buffer
Separation Buffer
Buffer Tray
Droppers
0.2 ml tubes

6. Click “Change Buffer”.

7. Place Alignment Marker 20 bp &

1000 bp C109100 (≥ 20 μ

l) at MA-1 

position and the buffer tray on the 
tray holder.
For other size of Alignment Marker, 
please place to the corresponding 
position.

* Please make sure buffer tray have 

been placed correctly and fitted in 
the holder.

* Spin down and make sure the 

Alignment Marker is in the bottom 
of tube and no air bubble appears. 

8. Hold the tray holder by your hand 

and press down the tube of 
Alignment Marker tightly into the 
well by thumb.

9. Click “Change Sample” and place your samples. (≥ 20 μ

l)

* Spin down and make sure the sample is in the bottom  

of tube and no air bubble appears.

10. Open the cartridge door and insert 

cartridge. 
(Guiding groove facing front)

11. Close the cartridge door.

* Press the door and make sure 

you hear the squeeze sound .

12. Click “Latch”.

Cartridge information will be presented on the main 
window after the cartridge is latched.

Bio

Fragment Analyzer :

Qsep

100

Document No.: F0004
Ver.: E-2021

Page: 1/3

(P)

(W)

(S)

(C)