18
2.3 Hardware Installation
Place
Qsep
™
Series instrument on a stable platform, and ensure it is away from
water and any other high-power electronic devices
.
Qsep
™
Series need to be
operated in a clean and well-ventilated environment. Remove the power adapter
from the package and attach it with a suitable power cord.
When using the portable air pump as the air source, please pay attention to the
condensation water that might develop inside the portable air pump. If this occurs,
be sure to release the condensation water before using
Qsep
™
Series. If you need to
continuously operate
Qsep
™
Series, check the air pump every 12 hours to avoid the
mist of condensation water damaging the devices of
Qsep
™
Series.
2.3.1
Qsep
100
™ Installation
Prepare a computer with Windows operation system (i.e. Windows 10). Use the
supplied USB cable to connect
Qsep
100
™
with the computer. Turn ON the power of
the computer and
Qsep
100
™
and install the operation software,
Q-Analyzer
™.
After checking all connections, you are now ready to start using
Qsep
100
™.
Figure 2-1
Qsep
100
TM
Hardware installation
Содержание Qsep1
Страница 9: ...9 Figure 1 2 Rotation platform...
Страница 11: ...11 1 1 3 Qsep100 Top View Figure 1 4 Qsep100 top view A Cartridge door...
Страница 14: ...14 1 2 3 Qsep1 Top View A Cartridge door A...
Страница 34: ...34 Introduction of the Control Panel...