Cautions

 

Qsep

™ 

Series

 

are capillary electrophoresis instruments driven by high voltage.   

Carefully read and thoroughly comply with the following instructions to maintain the 

integrity of the equipment, the reliability of results, and the safety of operation.   
Keep 

Qsep

™ 

Series away from other electronic device and voltage sources.

 

 

1.

 

Only  the  components  and  consumables  provided  by  BiOptic  Inc.  are 

suggested to use.

 

2.

 

DO NOT

 perform the following actions:

 

 

Disassemble 

Qsep

™ Series instrument and its associated parts

 

 

Bumps or jolts to 

Qsep

™ Series 

 

 

Move  the  instrument or  remove  any connected external  equipment, 
such as computers and air pumps, while 

Qsep

™ Series are in action

 

 

Open  the  sample  door  or  cartridge  door  while 

Qsep

™  Series  are  in 

action

 

 

Remove the cartridge while 

Qsep

™ Series are in action

 

 

Power off 

Qsep

™ Series before confirming 

Qsep

™ Series are inactived

 

3.

 

Store the cartridge upright; do not lay the cartridge on a table horizontally 

or place it upside down.

 

4.

 

Store the cartridge in the calm-shell container (Cartridge needs to be stored 

in a proper environment as the instructions of cartridge unpacking guide to 

maintain its lifetime and quality).

 

5.

 

Ensure  the  cartridge  is  taking  out  from

  Qsep

400

™  before  shutting  down  or 

disconnecting instrument. Or ensure the cartridge is placed at 

PARK

 position 

before  shutting  down or  disconnecting 

Qsep

™  Series.  Failure to do  so  will 

cause dryness of the gel at the cartridge tip, which will damage the cartridge.

 

6.

 

Before operating 

Qsep

™ Series

ensure the buffer tray is placed properly. 

 

7.

 

Dilute  the  PCR  DNA  test  sample  to 

1

20

   with  our  dilution  buffer  before 

running the test for obtaining better result and retaining the lifetime of the 

cartridge. You can adjust the dilution ratio after primary test runs. 

 

8.

 

Ensure  that  the  test  sample,  alignment  marker,  cartridge,  and  buffer  tray 

have been removed from the instrument before transporting or shipping. 

 

9.

 

If the buffer solution spills out on the platform inside the instrument, wipe 
it with tissue and clean it up immediately.

 

10.

 

Ensure that the main power of the instrument has been shut down and all 

Содержание Qsep 400

Страница 1: ...BiOptic Inc 4F No 108 3 Minquan Rd Xindian Dist New Taipei City 23141 Taiwan R O C www bioptic com tw Revolution Qsep400 Operation Manual Hardware...

Страница 2: ...s 4 Packing List 4 Cautions 6 1 Qsep400 Instrument 9 1 1 Qsep400 Front View 9 1 2 Qsep400 Back View 11 1 3 Qsep400 Top View 12 2 Installation Instruction 13 2 1 Environment Requirements 13 2 2 Power C...

Страница 3: ...is prohibited without prior written permission of BiOptic Inc except as allowed under the copyright laws Qsep Series and Q Analyzer are registered trademarks of BiOptic Inc Microsoft Windows is regist...

Страница 4: ...INSTRUCTIONS BiOptic Inc assumes no liability whatsoever for any personal injury property damage or other loss resulting from not complying or familiar with the manuals or improper operation of the d...

Страница 5: ...allation Disc including Q Analyzer for Qsep400 and Qsep Series Operation Manual Note The Software key contains 8GB storage space Q Analyzer for Qsep400 and Qsep Series Operation Manual including Hardw...

Страница 6: ...ce it upside down 4 Store the cartridge in the calm shell container Cartridge needs to be stored in a proper environment as the instructions of cartridge unpacking guide to maintain its lifetime and q...

Страница 7: ...lease disconnect the power plug 15 Please clean the instrument with dry cloth For other maintenance please contact local distributor 16 Please use the MAINS power cord within adequate rating 17 The co...

Страница 8: ...ntuitively The disposable gel cartridge with integrated pre programmed test methods makes capillary electrophoresis experiments no longer a painful procedure that requires well trained operators No mo...

Страница 9: ...e high voltage source of the capillary electrophoresis is ON B Main Power switch The blue LED lights up when it is turned on C Operation room Open the onboard computer module A rotation platform inclu...

Страница 10: ...10 Figure 1 2 Rotation platform...

Страница 11: ...iew A Air pressure gauge Display the value of the pressure operating in Qsep400 B Air flow Inlet Port Connect to an external pump using an 1 8 inches tube C USB Port Connect to a computer D Power Conn...

Страница 12: ...12 1 3 Qsep400 Top View Figure 1 4 Qsep400 top view A Cartridge door...

Страница 13: ...installation it is recommended to leave approximately 711 mm 28 inches clear space above the counter space or table Allow approximately 76 mm 3 inches of clear space at the rear and to the side of the...

Страница 14: ...in the package 100 240V AC 47 63Hz The dimensions are 12 2cm L 5 2cm W 3 5cm H and weight of the instrument is 26Kg Direct control the instrument with onboard computer Connect the air source within th...

Страница 15: ...ace Qsep400 instrument on a stable platform and ensure it is away from water and any other high power electronic devices Qsep400 needs to be operated in a clean and well ventilated environment Remove...

Страница 16: ...Qsep400 The message box of Purge function check will pop up after connected Follow the instruction and proceed the check before using is recommended User needs to perform the purge function check ever...

Страница 17: ...y into the tray holder Step 6 Alignment Marker AM Preparation Add at least 20 l Alignment Marker into the 0 2 ml tube 4 tubes and add 10 l Mineral Oil on the top of Alignment Marker Spin down and make...

Страница 18: ...l tube that placed at A1 A12 position must be removed cutoff Step 10 Click Park to move the holder back to the park position and click Close Sample Door Step 11 After unpacking cartridge open the cart...

Страница 19: ...se proceed with the following steps The storage and transportation condition may influence the cartridge and cause unstable current If current gray line is unstable during HV check please repeat this...

Страница 20: ...lumn and designate the sample locations test method and result name following steps 14 1 to 14 4 Step 14 1 Sample Position Mark the position of sample on the plate and then click OK Step 14 2 Method S...

Страница 21: ...Para Set the parameters Baseline Factor Peak Threshold Calculate Calculate Reference marker table based on built in size marker data Create size marker Place size marker at SM row on the holder create...

Страница 22: ...the results During Calibration Check the alignment markers have been placed in the correct position Software will recognize two peaks from Alignment Marker signal DO NOT use Size Marker or DNA sample...

Отзывы: