15
2.3 Hardware Installation
Place
Qsep
™
Series instrument on a stable platform, and ensure it is away from water
and any other high-power electronic devices
.
Qsep
™
Series need to be operated in a
clean and well-ventilated environment. Remove the power adapter from the package
and attach it with a suitable power cord.
When using the portable air pump as the air source, please pay attention to the
condensation water that might develop inside the portable air pump. If this occurs, be
sure to release the condensation water before using
Qsep
™
Series. If you need to
continuously operate
Qsep
™
Series, check the air pump every 12 hours to avoid the
mist of condensation water damaging the devices of
Qsep
™
Series.
2.3.1
Qsep
400
™
Installation
Place
Qsep
400
™
instrument on a stable platform, and ensure it is away from water and
any other high power electronic devices.
Qsep
400
™
needs to be operated in a clean
and well ventilated environment. Remove the power adapter from the package and
attach it with a suitable power cord. Connect
Qsep
400
™
to a properly grounded power
plug or power supply. Connect the air source, within the range of 0.3Mpa~0.5Mpa
(45psi~75psi), to the air inlet port at the back of
Qsep
400
™
. Please ensure the quality
of air source and prevent the damage caused by humid and dirty air. An air pump with
filter is suggested to be the air supplier. Alternative, you can use N2 gas as an air
source.
In
Qsep
400
™
instrument, it is built-in the operation software,
Q-Analyzer
™
.
After checking all connections, you are now ready to use
Qsep
400
™
. Please follow the
steps of operation quick start to operate the electrophoresis analysis test.
Figure 2-1
Qsep
400
TM
Hardware installation