
13
2. Installation Instruction
A suitable operating environment is essential to ensuring the best performance of
Qsep
™
Series.
2.1 Environment Requirements
⚫
The optimal operating temperature of
Qsep
™ Series are 18°C
~
26°C (65°F
~
78°F).
⚫
The optimal operating humidity of
Qsep
™ Series
are 40%
~
95% RH, non-
condensing.
⚫
The operating altitude of
Qsep
™ Series
are -20m
~
2000m (-65ft
~
6500ft).
For proper capillary electrophoresis operation, do not place
Qsep
™
Series on a
vibrating surface or near vibrating objects. For convenient access for loading and
installation, it is recommended to leave approximately 711 mm (28 inches) clear space
above the counter space or table. Allow approximately 76 mm (3 inches) of clear space
at the rear and to the side of the instrument for proper air ventilation and electrical
connections.
Warning
: Do not store, transport, or use
Qsep
™
Series under conditions which the
temperature fluctuation could cause water condensation inside the instrument.
Condensation may damage electronic devices of the instrument. If the instrument was
transported in cold weather, do not remove the instrument from its original box
immediately because it might cause condensation. Make sure
Qsep
™
Series
have
been warmed up inside the original box before you start unpacking.