TROUBLESHOOTING

CHECK THIS SECTION BEFORE CALLING OR SENDING YOUR pH BOSS FOR SERVICE.

Symptom

Possible Causes and Remedies

Unstable or inaccurate 
pH readings

pH reading cannot be 
calibrated

pH reads around pH7 
without changing

pH reads around pH4 to 5 
without changing

Unit does not power up

Alarm sounds due to pH 

being too low

Acid Pump turns, but acid 

is not flowing.

Unit is not adding acid 
when dosing pump is 
operating

pH Probe is dirty.

!

Disconnect probe  BNC connector from the pH  BOSS, then unscrew probe 

from the pipe.

!

Wipe the glass bulb with a soft tissue soaked with methylated spirits.

!

Soak the tip of the probe in diluted Hydrochloric acid (for example, from your 

pH BOSS diluted acid drum). 

!

Rinse the pH Probe thoroughly in clean water.

!

Screw probe back into pipe and re-connect probe BNC connector.

pH Probe is dirty.

!

Follow the pH Probe cleaning procedure above.

Short circuit in pH Probe BNC Plug or Socket

!

Disconnect pH Probe plug from socket. Dry out with a hair dryer or allow to air 

dry. Reconnect and re-check.

!

If problem persists, return pH Probe to have connector replaced, or replace 

the pH probe. Please contact your distributor or Bionizer.

Glass bulb or internal stem of pH Probe is cracked.

!

Replace pH Probe. Please contact your distributor or Bionizer.

Faulty AC/DC power adaptor

!

Replace power adaptor. Please contact your distributor or Bionizer.

!

Faulty pH BOSS main unit. Return to Bionizer for repair.

Dose Setting is set too high

!

Reduce the Dose Setting by one level and check in 2 days.

Leaking or worn squeeze tube and/or rollers.

!

Tighten Acid Pump fittings further or replace with spare fittings (see page 8).

!

Replace squeeze tube and/or rollers as needed (see page 14).

!

Check that the pH BOSS diluted acid drum is not empty.

!

Check that all suction and output hoses are fitted correctly, with no leakage.

!

Check that squeeze tube inside pump is not leaking.

WARNING  :  IF  ANY  DILUTED  ACID  HAS  LEAKED  OUT  OF  HOSES,  WASH 
DOWN THE AREA WITH FRESH WATER IMMEDIATELY. TAKE THE LID OFF 
THE PUMP, REMOVE SQUEEZE TUBE AND WASH OUT WITH FRESH WATER 
IF SQUEEZE TUBE WAS LEAKING.

pH Probe is faulty.

!

Replace pH Probe. Please contact your distributor or Bionizer.

pH Probe needs to be re-calibrated.

!

Clean pH Probe as per procedure above.

!

Re-calibrate pH Probe as per details on page 12.

Pool water is very pure and has low conductivity.

!

Add one bag of pool salt per 50,000 Litres (13,000 Gallons) of pool water.

IF YOU NEED TO RETURN YOUR pH BOSS FOR SERVICE OR REPAIRS AT ANY TIME, PLEASE USE THE MANUAL DOSE 
FUNCTION (PAGE 10) TO DOSE 1 LITRE (1 QUART) OF FRESH WATER THROUGH THE SYSTEM BEFORE SENDING BACK. 
THIS WILL STOP DILUTED ACID LEAKING FROM THE PUMP DURING TRANSPORT (WHICH WOULD DAMAGE YOUR 
UNIT OR THE PACKAGING, VOIDING WARRANTY), AND IS ALSO A SAFETY HAZARD FOR OUR TECHNICIANS.

15

TROUBLESHOOTING

Содержание PH BOSS

Страница 1: ...TM INSTRUCTION BOOK Automatic Pool pH Control System Please read before you call us for advice Version 1 51d 07 Nov 2017 bionizer com au bionizer com DOWN UP AUTO OFF CANCEL PROGRAM SET TM...

Страница 2: ...PUMP HOSE INSTALLATION 8 TEE PIECE HOSE INSTALLATION 9 DRUM WEIGHT HOSE INSTALLATION 9 pH PROBE INSTALLATION 9 PROGRAMMING YOUR pH BOSS 10 AUTO OFF KEY 10 DOSE SETTING 10 MANUAL DOSE 11 CHANGING THE p...

Страница 3: ...uted acid every weekortwo So the pH BOSS saves you time frustration AND money Before installing the pH BOSS the pool water should be correctly balanced particularly total alkalinity The pH BOSS can be...

Страница 4: ...grammingmenus Presstoscrollthroughthefiveusersettingmenus Presstosavechangeswhenadjustingsettings OK pHreadingiscorrect asperSetPointsetting Add pHreadingisaboveSetPoint pH BOSS dosingpumpwillrun whil...

Страница 5: ...e pH BOSS from a genuine Bionizer power adaptor An incorrect adaptor could damage the pH BOSS and will void your warranty 1 WARNING alwaysusedilutedacidat1partacid to4ormorepartswaterwithpH BOSS 2 ALW...

Страница 6: ...installed fitthepH BOSS tee BEFORE the Salt Chlorinator cell The acid will help to keep the expensive Chlorinator cellsclean If a liquid Chlorine injection system is fitted install the tee BEFORE the...

Страница 7: ...mvertical DO NOT FIT THE TEE TO A VERTICAL PIPE AS THIS WILL MAKE THE pH PROBE HORIZONTAL WHICH WILL RENDER THE pH PROBE INOPERABLE Correctly Installed Tee Incorrectly Installed Tee Do not mount on ve...

Страница 8: ...easily Fix the control box in position usingscrewsateachcornerofthebox NOTE Make sure that the pH BOSS is positioned near enough to a power outlet and to the T piece and acid drum to allow the power c...

Страница 9: ...to the Tee piece thread Tighten firmly by hand DO NOT OVER TIGHTEN DO NOT USE A TOOL 3 After the pH Probe is installed in the Tee piece connect plug on the pH Probe cable into the probe socket on the...

Страница 10: ...terto1partacid page5 For pools with a wet edge and balance tank use a settingof1lowerthanlistedforthepoolsize If the pH BOSS does not keep up with acid demand and the Alarm sounds increase the dose se...

Страница 11: ...x Re calibratethepHProbe Seepage12 Problem pHProbeisbrokenorfaulty Fix If re calibration is not possible then replacethepHProbe Seepage9 A replacement pH Probe can be ordered by telephone or online at...

Страница 12: ...ationjar seepicture ontheright Fill the Buffer Jar to the 250mL mark with distilled water Filtered tap water canalsobeused Replace both the seal and thelid thenshakewellfor 1 minute to ensure that t h...

Страница 13: ...topH BOSS whendone Re connect the pool pump to the Bionizer power outlet or the power output to which it was connectedbeforestartingthisprocedure Switchyourpoolequipmentbackon If using a Bionizer ensu...

Страница 14: ...he pump rotor a little and place the squeeze tube between the pump body and the two rollers Push the squeeze tube and rotor into the pump body Ensure that the inlet and outlet are seated in theguideso...

Страница 15: ...er Faulty AC DC poweradaptor Replacepoweradaptor PleasecontactyourdistributororBionizer FaultypH BOSS mainunit ReturntoBionizerforrepair DoseSettingissettoohigh ReducetheDoseSettingbyonelevelandchecki...

Страница 16: ...dale Rd Suite 410 Scottsdale AZ 85254 USA Phone 1 480 696 2387 Toll Free 1 855 8 IONIZE Email relax bionizer com Web bionizer com Bioniser Pty Ltd 1 173 Bradman Ave Maroochydore QLD 4558 Australia Pho...

Отзывы: