Thank you for selecting the 

Rightest

 

Blood Glucose Monitoring System GM232C. This manual provides all the 

information you need to operate this product for accurate test results. Please read this entire manual before you 

start any testing.

For people living with diabetes, it is important to regularly monitor blood glucose levels to effectively reduce 

complications from the disease. The easy-to-use 

Rightest

 

Blood Glucose Monitoring System GM232C provides 

accurate, reliable test results to help you better manage your diabetes. 

The 

Rightest

 

Blood Glucose Monitoring System GM232C is designed for in vitro diagnostic use only (for self-testing by 

a single user outside the body). Testing requires only a small amount of fresh capillary whole blood from either the 

fingertip, palm or forearm. 
The 

Rightest

 Blood Glucose Monitoring System GM232C is manufactured and supported by Bionime Corporation. If 

you have any questions or concerns, please contact the Bionime Customer Service toll free at 888-481-8485. We will 

make every effort to assist you.

 

1

Preface

Copyright © 2014 Bionime Corporation. All Rights Reserved Worldwide

Содержание GM232C

Страница 1: ...Sec 2 Daqing St South Dist Taichung City 40242 Taiwan R O C Made in Taiwan USA Service Center BIONIME USA CORPORATION 1450 E Spruce Street Bldg B Ontario CA 91761 USA Tel 888 481 8485 www gediabetes c...

Страница 2: ......

Страница 3: ...curate reliable test results to help you better manage your diabetes The Rightest Blood Glucose Monitoring System GM232C is designed for in vitro diagnostic use only for self testing by a single user...

Страница 4: ...be used to screen for or diagnose diabetes mellitus If the Rightest Blood Glucose Meter GM232C and Test Strips are exposed to a substantial change in temperature please wait 45 minutes before measure...

Страница 5: ...single patient use Do not share them with anyone including other family members All parts of the kit are considered biohazardous and can potentially transmit infectious diseases even after following...

Страница 6: ...View Window Appearance Understanding Test Results and Messages Rightest Rightest Rightest Rightest Quality Control About Quality Control Testing Performing a Quality Control Test Understanding Contro...

Страница 7: ...Control Solution GC550 Package Insert 5 Blood Glucose Meter GM232C with 2 AAA batteries installed 6 Blood Glucose Test strips GM232C 10 25 pcs 7 8 Lancing Device 9 Clear Cap 10 Disposable Sterile Lan...

Страница 8: ...on Main Button Display Window Test Strip Port Right Button Product name Battery Cover Meter Serial Number Meter Battery The Rightest Blood Glucose Meter GM232C The Rightest Blood Glucose Meter GM232C...

Страница 9: ...into meter Appears when an error occurs Current date under time mode or testing date under memory mode Warns when the operational temperature limit is exceeded during testing Indicates the time in 12...

Страница 10: ...ood Glucose Test Strips When you open a new vial of Blood Glucose Test Strips record the date on the vial Discard the vial of test strips after 3 months from opening Store the Blood Glucose Test Strip...

Страница 11: ...the local regulation to properly discard a used battery Risk of explosion if battery is replaced by an incorrect type 4 The Blood Glucose Meter 5 Press any button to exit the self test and enter Setti...

Страница 12: ...r the Setting Mode Press the main button to turn on the Rightest Blood Glucose Meter GM232C Hold down the main button for 7 seconds During this time the screen will go blank until you hear a beep Afte...

Страница 13: ...ress the Main button to confirm Once confirmed the Hour Setting will appear Once the hour is confirmed the Minute Setting will appear Once the minute is confirmed the Volume Setting will appear 5 6 4a...

Страница 14: ...n the screen 2 Turn off the Bluetooth In Time mode press and hold the Right or Left button for two seconds The blinking symbol will switch to the symbol with a beep If Volume is turned on 3 Manual tur...

Страница 15: ...ip port until it clicks and firmly stops Removing the Rightest Blood Glucose Test Strip GM232C 1 Hold the Rightest Blood Glucose Test Strip GM232C as shown 2 Rotate the Blood Glucose Test Strip counte...

Страница 16: ...justable depth cap in one hand while holding the base in the other hand Lift the cap until a gap appears between the cap and base 2 Pull the two parts in opposite directions to remove the adjustable d...

Страница 17: ...GM232C GM232C 7 8 NOTE The will automatically detect the Code number on the strip You do not have to check the Code number on the meter display and strip vial Rightest Blood Glucose Meter GM232C 12 A...

Страница 18: ...nd if you apply blood too soon If this occurs please repeat the test with a new test strip Record the opening date of a new test strip vial Discard the vial of test strips 3 months after opening Alway...

Страница 19: ...ge 45 for more information NOTE If no button pressed the marker will be set blank automatically after 15 seconds Pressing for a few seconds Releasing button Alternative site testing palm or forearm bl...

Страница 20: ...to a laboratory test system that is known to be well maintained and monitored by a healthcare provider use the clear cap provided with the Rightest GD500 Lancing Device when testing Make sure your bl...

Страница 21: ...immediately If you are experiencing symptoms that are not consistent with your blood glucose test results and you have made sure to follow all instructions in this manual contact your healthcare profe...

Страница 22: ...nt to purchase or another control solutions please contact the GE Diabetes Customer Support Center Rightest Control Solution GC550 Level 4 Level 2 toll free at 1 866 613 7085 When should a Quality Con...

Страница 23: ...ghtest Blood Glucose Test Strip from the vial and close the vial cap immediately 2 Insert the Blood Glucose Test Strip horizontally not vertically with the colored view window facing up into the test...

Страница 24: ...pted will result in an error message and Do not drip the control solution on to the sample port of the test strip directly The reagent on the strip might leak into the bottle of control solution and m...

Страница 25: ...appears under the data the test was performed using the The control solution test result will not be used when recalling test average results Rightest Control Solution GC550 1 Press the Main button t...

Страница 26: ...ve to MEM Mode Average Mode or Time Mode you will not be able to perform the exclusion The Lo Hi results control solution test results and test results under abnormal temperature 50 F 10 C 104 F 40 C...

Страница 27: ...le pathogens The meter and lancing device should be cleaned before being disinfected CaviWipes Disinfecting Towelettes has been tested and may be used to clean and disinfect the meter and lancing devi...

Страница 28: ...g cycles per week over the typical life of the meter 5 years 2 Allow to remain wet for 2 minutes 1 CAUTION Users should wash their hands thoroughly with soap and water after handling the meter lancing...

Страница 29: ...over 111 F 44 C the meter will not function and the symbol will blink 3 If the Rightest Blood Glucose Meter and Test Strips are exposed to a substantial change in temperature please wait 45 minutes be...

Страница 30: ...cover and remove the batteries 2 Wait for 5 minutes and reinsert the batteries as described as instructed on page 16 Meter Activation and Changing the Battery The meter should work normally after fin...

Страница 31: ...essive water loss may cause inaccurately low results High concentrations of Ascorbic acid Vitamin C 5 mg dL Cholesterol 600mg dL and Uric acid 10 mg dL may result in inaccurate test results Not for ne...

Страница 32: ...5 b This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC rules These limits are designed to provide reasonable protection agai...

Страница 33: ...600 mg dL 5 seconds 50 104 F 10 40 C 10 90 500 blood glucose test results with date and time Turns off automatically 2 minutes after last user action To turn off manually press themain button for 3 se...

Страница 34: ...dels are not covered with this warranty card GE is a trademark of General Electric Company and is used under license by Bionime Corporation No 100 Sec 2 Daqing St South Dist Taichung City 40242 Taiwan...

Страница 35: ...tructions to make sure you are performing the steps correctly If you have any questions or concerns please contact the Bionime customer service toll free at 888 481 8485 We will make every effort to p...

Страница 36: ...thout Diabetes References 1 American Diabetes Association Standard of Medical Care in Diabetes 2014 Diabetes Care Vol 37 supplement 1 S14 S80 January 2014 Expected glucose values without diabetes Stat...

Страница 37: ...Klaus Ellensohn Tschuetschgasse 8 6833 Klaus Vlbg Austria Disposable Sterile Lancets Manufacturer SteriLance Medical SuZhou Inc No 68 LiTangHe RD XiangCheng SuZhou JiangSu 215133 P R China Product co...

Страница 38: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 39: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 40: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 41: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 42: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 43: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 44: ...n Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Breakfast Lunch Dinner Bedtime Other Comments DATE M T W T F S S Blood Glucose nsulin Me...

Страница 45: ...omments DATE M T W T F S S Blood Glucose nsulin Medication I Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Blood Glucose Insulin Medication Warrant...

Страница 46: ...your Blood Glucose Meter GM232C 3 The sole obligation of Bionime under this warranty shall be to replace any defective Blood Glucose Meter GM232C No other warranties express or implied are made Bionim...