39
38
About Quality Control Testing
About Quality Control Testing
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the
Rightest
Control Solution GC550 when testing your
Rightest
Blood Glucose Monitoring System
GM232C in the Control Solution Mode. If the test result is within the Control Solution Range printed on the strip vial
label, the
Blood Glucose Monitoring System GM232C passes the Quality Control Test and your
Blood Glucose Monitoring System GM232C is working properly.
Rightest
Rightest
Control Solution Range:
Example of Control Solution Range printed on
your test strip vial label.
Control Solution Range mg/dL
Level 2: 107-145
Level 4: 315-426
NOTE
- Level 2 of
is provided in your starter kit. If you want to purchase
or
another
control solutions, please contact the GE Diabetes Customer Support Center
.
Rightest
Control Solution GC550
Level 4
Level 2
toll free at
1-866-613-7085
When should a Quality Control Test Be Performed ?
- To ensure that your
Blood Glucose Meter
and
Blood Glucose Test Strip
are
working properly.
- To confirm that you are following the correct testing procedures.
- To prepare for your initial blood glucose test.
- To check the
Blood Glucose Test Strip
when you open a new vial of strips.
-
- To practice testing.
Required Items for Quality Control Tests
To perform a quality control test, prepare the items below :
-
Blood Glucose Meter
-
Blood Glucose Test Strips
-
Rightest
GM232C
Rightest
GM232C
Rightest
GM232C
To check your
Rightest
Blood Glucose Meter GM232C after it has been dropped, damaged or exposed to liquids.
- If you suspect that your test results are inaccurate, or if your test results are not consistent with the way you feel.
Rightest
GM232C
Rightest
GM232C
Rightest
Control Solution GC550
Содержание GM232C
Страница 2: ......