41
BM5Vet Pro
User’s Manual
3. Network
Overview
When you connect the monitor to the network, you can access animal information from another
monitor or central station connected to the network. These devices provide main screen
information for remote viewing from each other.
BM Vet Central System(BT-Link, B-Link) connects the monitors to the central station and each
device to provide various monitoring functions.(For more information about the central station,
see the BM Vet Central (BT-Link, B-Link) Station User Guide.)
The User Monitor's B2B View (Bed to Bed View) feature allows the user to view other monitor
screens connected to the network and to silence remote control and alarms [Audio Paused].
With the Remote Control feature in BM Vet Central station (BT-Link, B-Link), you can perform the
following tasks on a veterinary patient monitor that can be remotely controlled from a central
station.
⚫
Start recording
⚫
Modify alarm limit
⚫
Alarm Mute
⚫
Start to check Arrhythmia
⚫
Enter, edit and view animal data
Network connection
In the network, data can be exchanged via wired or wireless technology. This device can exchange
information with other devices over the network during operation and supports the following
features.
⚫
Display waveform and parameter data
⚫
Alarm signal
⚫
Remote control (e.g. alarm management)
⚫
View another Veterinary patient monitor by remote access
⚫
Device setup and transmission of animal data
Connecting this device to an integrated network with other devices, or subsequent changes to
that network, can be a new risk to animals, users, and third parties. These risks must be identified,
analyzed and evaluated before the device is connected to the network or the network is changed,
Содержание BM5VET
Страница 22: ...22 BM5Vet Pro User s Manual Left Side 1 Printer...
Страница 30: ...30 BM5Vet Pro User s Manual How To Replace The Battery Please assemble and replace as shown below...
Страница 68: ...68 BM5Vet Pro User s Manual ECG Lead 5 Positions of 5 Lead Placement RL V LL LA RA...
Страница 155: ...155 BM5Vet Pro User s Manual Accessories Bionet Dual Gas Module...
Страница 195: ...195 BM5Vet Pro User s Manual ACCVENT COUPLET IRREGULAR PAUSE RONT VBRADY SHORTRUN PVC...
Страница 219: ...219 BM5Vet Pro User s Manual SYSTEM LOWBATTERY...
Страница 225: ...225 BM5Vet Pro User s Manual less than minus number percent plus or minus...